Owner's manual

<2. Connection>
2-2
IM 01C25R02-01E
(1) Conrmation by the name plate
The HART protocol revision is shown by the last
number of the serial number.
Refer to Figure 2.1.
F0201.ai
: Refer to USER'S MANUAL.
Made in Japan
TOKYO 180-8750 JAPAN
MODEL
SUFFIX
SUPPLY
OUTPUT
MWP
mA DC
V DC
STYLE
CAL
RNG
NO.
XXX - - - XX
5 : HART 5
7 : HART 7
Figure 2.1 Name Plate
(2) Conrmation by using integral indicator (When
the integral indicator code is “D”)
Refer to section 2.1.
(3) Conrmation by using HART conguration tool
Follow the procedure below.
1) Connect the conguration tool to the
transmitter.
2)-1 When using DD(HART 5/HART 7) or
DTM(HART 7) conguration tool.
Select the “Root Menu(Refer to
subsection 3.1.1)
Call up the “Reviewdisplay.
2)-2 When using DTM(HART 5) conguration
tool.
Call up “HART” display.
Conguration HART
3) HART protocol revision is displayed on the
“Universal revcolumn.
IMPORTANT
Protocol revision supported by HART
conguration tool must be the same or higher
than that of the EJX multivariable transmitter.
Protcol revision
supported by HART
conguration tool
5 7
Protocol revision of
EJX multivariable
transmitter
5
7 ×
: Communication OK
×: Communication NG
NOTE
HART 7 communication is supported by
FieldMate R2.02 or later.
NOTE
When the output signal code of EJX multivariable
transmitter is “-J”, HART protocol revision can be
changed. Refer to subsection 3.3.17 about the
procedure of the revision change of HART 5 and
HART 7.
2.3 Device Description (DD) on
a Conguration Tool and
Transmitter Device Revision
Before using a HART conguration tool, conrm
that the DD for the transmitter is installed in the
conguration tool.
About the DD, use the device type, device revision
and DD Revision shown in the Table 2.1.
Table 2.1 HART Protocol Revision, Device
Revision and DD Revision
HART
Protocol
Revision
(*1)
EJX Multivariable Transmitter
Model
Name
Device
Type
Device
Revision
DD
Revision
5
EJX910A
EJX930A
EJX-MV
(0x54)
2
2
or later
7
EJX910A
EJX930A
EJX-MV
(0x3754)
10
1
or later
*1: When the output signal code is “–E”, only “5” is available.
The device revision of the transmitter and DD can
be conrmed as shown below.
If the correct DD is not installed in the conguration
tool, download it from the ofcial web site of HART