Owner's manual

<3. Parameter Setting>
3-21
IM 01C25R02-01E
(1) Auto Sensor Trim
Applying reference pressure of 0% and 100% of the
measurement range to the transmitter, adjust the
lower and upper points automatically.
Procedure to call up the display
DD (HART 5/7)
DTM (HART 7)
[Root Menu] Diag/Service
Calibration Pres sensor trim
Pres trim
DTM (HART 5) Calibration Pressure trim
Auto, Lower Pt Auto trim for 0% point
Auto, Upper Pt Auto trim for 100% point
(2) Manual Sensor Trim
Using the example below, follow the steps to
perform the full sensor trim by manually. The Pres
LTD (Manual, Lower Pt) and Pres UTD (Manual,
Upper Pt) represent the previously adjusted values.
Example: For the range of 1000 to 3000 mmH
2
O
Pres LTD (Manual, Lower Pt) = 4.0 mmH
2
O
Pres UTD (Manual, Upper Pt) = −3.0 mmH
2
O
<1> Call up the Manual, Lower Pt.
Procedure to call up the display
DD (HART 5/7)
DTM (HART 7)
[Root Menu] Diag/Service
Calibration Pres sensor trim
Pres trim
DTM (HART 5) Calibration Pressure trim
Manual, Lower
Pt
Manual trim for 0% point
Manual, Upper
Pt
Manual trim for 100% point
<2> Suppose that a standard pressure of 1000
mmH
2
O is applied and the value of the “Pres
for trim” is 994.0. Correct for this output error
of 6 mmH
2
O by adding 6 mmH
2
O to Pres LTD
(Manual, Lower Pt).
−4.0+6.0=+2.0
<3> Enter the correction value of “2” to the Pres
LTD (Manual, Lower Pt).
<4> Call up the Pres UTD (Manual, Upper Pt).
<5> Suppose that a standard pressure of 3000
mmH
2
O is applied and the value of the Pres for
trim is 3015.0. Firstly, obtain the slope error for
the span as follows;
Slope Error =
Applied Pressure ValueValue of Pres for Trim
× (URV−LRV)
Applied Pressure Value
=
3000−3015
× (3000−1000) = −10
3000
Then correct for this slope error of −10 by
adding −10 to Pres UTD (Manual, Upper Pt).
−3.0+(−10.0)=−13.0
<6> Enter the correction value of “−13to the Pres
UTD (Manual, Upper Pt).
(3) Sensor Trim for Static Pressure or External
Temperature
For the EJX multivariable transmitter, full sensor
trim of the static pressure or external temperature is
performed in the same way as with the differential
pressure.
Procedure to call up the display for static
pressure
DD (HART 5/7)
DTM (HART 7)
[Root Menu] Diag/Service
Calibration SP sensor
trim SP trim
DTM (HART 5) Calibration Static Pressure
trim
Auto, Lower Pt Auto trim for 0% point
Auto, Upper Pt Auto trim for 100% point
Manual, Lower Pt Manual trim for 0% point
Manual, Upper Pt Manual trim for 100% point
Procedure to call up the display for external
temperature
DD (HART 5/7)
DTM (HART 7)
[Root Menu] Diag/Service
Calibration ET Sensor
trim ET trim
DTM (HART 5) Calibration External Temp
trim
Auto, Lower Pt Auto trim for 0% point
Auto, Upper Pt Auto trim for 100% point
Manual, Lower Pt Manual trim for 0% point
Manual, Upper Pt Manual trim for 100% point
(4) Reset Trim Adjistment to Factory Setting
The Clear P trim, Clear SP trim and Clear ET trim
commands can reset the trim adjustment to the
initial calibrated values that were set. The amount of
the adjustment performed with the external zero-
adjustment screw is returned to the initial setting as
well.