    <10.  General Specications>
10-4
IM 01C25R01-01E
  Mass Flow Calculation
(For Measurement function code B of HART and 
FOUNDATION Fieldbus protocol types)
 
Auto Compensation Mode (FlowNavigator 
is required for conguration)
Conguration of the uid physical properties and 
primary device for the EJX910A/EJX930A can be 
performed using a dialog window of FlowNavigator.
All ow factors for mass ow calculation are 
dynamically compensated to an optimum value.
In Auto mode, mass ow can be measured with 
high accuracy.
The ow factors automatically compensated are 
discharge coefcient, diameter of primary device, 
upstream internal pipe diameter, gas expansion 
factor, density, and viscosity.
FlowNavigator: (Refer to GS 01C25R51-01E) 
FSA120 software package is used to perform mass 
ow conguration for the EJX910A/EJX930A. 
This software can also read and write the general 
transmitter parameters. Conguration of the uid 
physical properties and the primary device of the 
EJX910A/EJX930A can be done by means of a 
dialog menu.
FlowNavigator runs on a notebook PC equipped 
with a communication interface for HART/
F
OUNDATION Fieldbus.
  Basic Flow Calculation Mode
Flow operation and density compensation are 
performed conventionally, with the ow factors 
manually input.
The operational expression is switched by the uid 
type and the unit setting.
Density is compensated as follows according to the 
selection of gas or liquid
Gas: Compensation as ideal gas by 
temperature and pressure.
Liquid: Compensation by temperature.
Flow operational expression: refer to Table 2.1
Symbol: refer to Table 2.2
Flow unit categories: refer to Table 2.3, 2.4, 2.5
  Normal Operating Condition:
(Selected features may affect limits.)
Ambient Temperature Limits
–40 to 85°C (–40 to 185°F)
–30 to 80°C (–22 to 176°F) with LCD display
Process Temperature Limits
–40 to 120°C (–40 to 248°F)
Ambient Humidity Limits
0 to 100% RH
Working Pressure Limits (Silicone oil)
Maximum Pressure Limits
EJX910A
L Capsule 16 MPa (2300 psi)
M and H Capsule 25 MPa (3600 psi)
EJX930A
M and H Capsule 32 MPa (4500 psi)
Minimum Pressure Limit
See graph below
Atmospheric
pressure
-40
(-40)
0
(32)
40
(104)
80
(176)
120
(248)
1(0.14)
2.7(0.38)
10(1.4)
(psi abs)
100(14.5)
Process temperature °C (°F)
Working
pressure
kPa abs
Applicable range
F1002.ai
Figure 10.1 
Working Pressure and Process Temperature