Cause
4-20mA Output  
operation during error  
Countermeasure  
Status 
group  
AL. 41 F.HI   F high alarm   Input ow exceeds specied 
threshold.  
Continues to operate and 
output.  
Check input.  
5  
AL. 42 F.LO   F low alarm  
AL. 35 P.HI   P high alarm   Input pressure exceeds 
specied threshold. 
AL. 36 P.LO   P low alarm  
AL. 37 SP.HI   SP high alarm   Input static pressure 
exceeds specied 
threshold. 
AL. 38 SP.LO   SP low alarm  
AL. 43 ET.HI   ET high alarm   Input external temperature 
exceeds specied 
threshold. 
AL. 44 ET.LO   ET low alarm  
AL. 50  
P. LRV 
Illegal P LRV   Specied value is outside of 
setting range.  
Holds at the output value that 
existed immediately before the 
error occurred.  
Check settings and 
change them as 
needed.  
6
AL. 51  
P. URV 
Illegal P URV  
AL. 52  
P. SPN  
Illegal P SPAN  
AL. 53  
P. ADJ  
P SPAN trim err   Continues to operate and 
output.  
Adjust settings and 
change them as 
needed.
P ZERO trim err  
AL. 54  
SP. RNG  
Illegal SP LRV   Holds at the output value that 
existed immediately before the 
error occurred.  
Check settings and 
change them as 
needed.  
Illegal SP URV  
Illegal SP SPAN  
AL. 55  
SP. ADJ  
SP SPAN trim err   Continues to operate and 
output.  
Adjust settings and 
change them as 
needed.  
7 
SP ZERO trim err  
AL. 56  
ET. RNG  
Illegal ET LRV Holds at the output value that 
existed immediately before the 
error occurred.  
Check settings and 
change them as 
needed.  
8
Illegal ET URV
Illegal ET SPAN
AL. 57  
ET. ADJ  
ET SPAN trim err   Continues to operate and 
output.  
Adjust settings and 
change them as 
needed.  
7 
ET ZERO trim err  
AL. 58  
FL. ADJ  
F set outside Range   Specied value is outside of 
setting range.  
Holds at the output value that 
existed immediately before the 
error occurred.  
Check settings and 
change them as 
needed.  
8
AL. 59  
PLS.ADJ  
PLS set err   Specied value is outside of 
setting pulse output.  
Normal calculation.  
7
AL. 79  
OV. DISP  
(None)   Displayed value exceeds 
limit.  
Continues to operate and 
output.  
— 