Manual
    <9.  Maintenance>
9-4
IM 01C25R01-01E
9.4  Disassembly and 
Reassembly
This section describes procedures for disassembly 
and reassembly for maintenance and component 
replacement.
Always turn OFF power and shut off and release 
pressures before disassembly. Use proper tools for 
all operations. Table 9.2 shows the tools required.
Table 9.2   Tools for Disassembly and Reassembly
Tool Quantity Remarks
Phillips 
screwdriver
1 JIS B4633, No. 2
Slotted 
screwdriver
1
Allen wrenches 3 JIS B4648 
One each, nominal 3, 4 and 
2.5 mm Allen wrenches
Wrench 1 Width across ats, 17 mm
Torque wrench 1
Adjustable 
wrench
1
Socket wrench 1 Width across ats, 16 mm
Socket driver 1 Width across ats, 5.5 mm
Tweezers 1
CAUTION
Precautions for ATEX Flameproof Type 
Transmitters
•  Flameproof type transmitters must be, as 
a rule, removed to a non-hazardous area 
for maintenance and be disassembled and 
reassembled to the original state.
•  On the ameproof type transmitters the two 
covers are locked, each by an Allen head 
bolt (shrouding bolt). When a shrouding bolt 
is driven clockwise by an Allen wrench, it is 
going in and cover lock is released, and then 
the cover can be opened.
  When a cover is closed it should be locked 
by a shrouding bolt without fail. Tighten the 
shrouding bolt to a torque of 0.7 N·m.
F0903.ai
Shrouding Bolt
Shrouding Bolt
Figure 9.3  Shrouding Bolts
9.4.1  Replacing the Integral Indicator
CAUTION
Cautions for Flameproof Type Transmitters
Users are prohibited by law from modifying the 
construction of a ameproof type transmitter. 
This would invalidate the agency approval for the 
use of the transmitter in a rated area.
It follows that the user is prohibited from using 
a ameproof type transmitter with its integral 
indicator removed, or from adding an integral 
indicator to a transmitter. If such modication is 
absolutely required, contact Yokogawa.
This subsection describes the procedure for 
replacing an integral indicator. (See gure 9.4)
■  Removing the Integral Indicator
1)  Remove the cover.
2)  While supporting the integral indicator with one 
hand, loosen its two mounting screws.
3)  Dismount the LCD board assembly from the 
CPU assembly.
  When doing this, carefully pull the LCD board 
assembly straight forward so as not to damage 
the connector pins between it and the CPU 
assembly.
■  Attaching the Integral Indicator
1)  Align both the LCD board assembly and CPU 
assembly connectors and engage them.
2)  Insert and tighten the two mounting screws.
3)  Replace the cover.
F0904.ai
Press
forward
Output terminal cable
Zero-adjustment
screw pin
Slide 
switch
Boss
Integral
indicator
Mounting screw
Amplifier Cover
CPU assembly
Bracket
(for zero-adjustment
screw pin)
LCD board assembly
Zero-
adjustment
screw
Figure 9.4  Removing and Attaching LCD Board 
Assembly and CPU Assembly (HART 