    <9.  Maintenance>
9-3
IM 01C25R01-01E
V
RRc
P
Using pressure
generator
Reference pressure
High pressure side High pressure side
Low pressure side
open to atmosphere
Low pressure side
open to atmosphere
Supply pressure
Pressure generator
Load resistance,
250Ω
Load 
adjusting 
resistance, 
100Ω
Digital voltmeter
Power 
supply 
E
V
RRc
P
Using pressure
source with
manometer
Reference pressure
Pressure source
Model MT220 
precision digital manometer
Load resistance,
250Ω
Load 
adjusting 
resistance, 
100Ω
Digital voltmeter
Power 
supply 
E
F0901.ai
Figure 9.1  Instrument Connections for Differential Pressure (HART protocol type)
V
RRc
P
Using pressure
generator
Reference pressure
High pressure side High pressure side
Low pressure 
side
Low pressure 
side
Supply pressure
Pressure generator
Load resistance,
250Ω
Load 
adjusting 
resistance, 
100Ω
Digital voltmeter
Power 
supply 
E
V
RRc
P
Using pressure
source with
manometer
Reference pressure
Pressure source
Model MT220 
precision digital manometer
Load resistance,
250Ω
Load 
adjusting 
resistance, 
100Ω
Digital voltmeter
Power 
supply 
E
F0902.ai
Figure 9.2  Instrument Connections for Static Pressure (HART protocol type)
NOTE
Apply pressure to H side and L side evenly.