9.3.2  External Temperature (RTD)
Using a thermometer resistor as input, calibration 
of the temperature transmitter is carried out via a 
3-core wire connection.
As dened the reference resistor value table of 
the thermometer resistor (RTD), obtain resistance 
values corresponding to 0, 25, 50, 75 or 100% 
of the span, and use the obtained resistance as 
the input value, then deliver it to the temperature 
transmitter by means of a variable resistor. Measure 
the resulting output signal with the voltmeter (digital 
multimeter) and check the output value relative to 
the input value.
If the output signal deviates from the given range 
of accuracy when a given input signal is delivered, 
adjust the output using the handheld terminal. 
For details of how to adjust the output, refer to 
the additional references, “HART Protocol” IM 
01C25R02-01E and the instruction manual for each 
terminal.
Table 9.1  Instruments Required for Calibration
Name Yokogawa-recommended Instrument Remarks
Power supply Model SDBT or SDBS distributor 4 to 20 mA DC signal
Load resistor Model 2792 standard resistor [250 Ω ±0.005%, 3 W]
Load adjustment resistor [100 Ω ±1%, 1 W]
Voltmeter Model 2501 A digital multimeter
Accuracy (10V DC range): ±(0.002% of rdg + 1 dgt)
Digital 
manometer
Model MT220 precision digital manometer
1) For 10 kPa class
  Accuracy: ±(0.015% of rdg + 0.015% of F.S.). . . . for 0 to 10 kPa
    ±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -10 to 0 kPa
2) For 130 kPa class
  Accuracy: ±0.02% of rdg. . . . . . . . . . . . . . . . . . . . for 25 to 130 kPa
    ±5 digits . . . . . . . . . . . . . . . . . . . . . . . . for 0 to 25 kPa
    ±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa
3) For 700 kPa class
  Accuracy: ±(0.02% of rdg + 3 digits) . . . . . . . . . . . for 100 to 700 kPa
    ±5 digits . . . . . . . . . . . . . . . . . . . . . . . . for 0 to 100 kPa
    ±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa
4) For 3000 kPa class
  Accuracy:  ±(0.02% of rdg + 10 digits). . . . . . . . . . for 0 to 3000 kPa
    ±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa
5) For 130 kPa abs class
  Accuracy: ±(0.03% of rdg + 6 digits) . . . . . . . . . . . for 0 to 130 kPa abs
Select a manometer having 
a pressure range close to 
that of the transmitter.
Pressure 
generator
Model 7674 pneumatic pressure standard for 200 kPa {2 kgf/cm
2
}, 25 kPa 
{2500 mmH
2O}
Accuracy: ±0.05% of F.S.
Requires air pressure 
supply.
Dead weight gauge tester 25 kPa {2500 mmH
2O}
Accuracy: ±0.03% of setting
Select the one having a 
pressure range close to that 
of the transmitter.
Pressure 
source
Model 6919 pressure regulator (pressure pump)
Pressure range: 0 to 133 kPa {1000 mmHg}
Prepare the vacuum pump 
for negative pressure 
ranges.
Variable 
resistor
279301 type 6-dial variable resistor (accuracy: ±0.005%) For calibration of 
thermometer resistor (RTD) 
input
Note:  The above table contains the instruments capable of performing calibration to the 0.2% level. Since special maintenance and 
management procedures involving traceability of each instrument to higher-level standards are required for calibration to the 0.1% 
or higher level, there may be difculties in calibration to this level in the eld. For calibration to the 0.1% level, contact Yokogawa 
representatives from which the instrument was purchased or the nearest Yokogawa ofce.