Manual
    <9.  Maintenance>
9-1
IM 01C25R01-01E
9.  Maintenance
9.1   Overview
WARNING
Since the accumulated process uid may be 
toxic or otherwise harmful, take appropriate care 
to avoid contact with the body or inhalation of 
vapors when draining condensate or venting gas 
from the transmitter pressure-detector section 
and even after dismounting the instrument from 
the process line for maintenance.
Maintenance of the transmitter is easy due to its 
modular construction. This chapter describes the 
procedures for calibration, adjustment, and the 
disassembly and reassembly procedures required 
for component replacement.
Transmitters are precision instruments. Please 
carefully and thoroughly read the following sections 
for information on how to properly handle them 
while performing maintenance.
IMPORTANT
•  As a rule, maintenance of this transmitter 
should be done in a shop that has all the 
necessary tools.
•  The CPU assembly contains sensitive parts 
that can be damaged by static electricity.
  Take precautions such as using a grounded 
wrist strap when handling electronic parts or 
touching the board circuit patterns. Also be 
sure to place the removed CPU assembly 
into a bag with an antistatic coating.
9.2  Calibration Instruments 
Selection
Table 9.1 lists the instruments that can be used 
to calibrate a transmitter. When selecting an 
instrument, consider the required accuracy level. 
Exercise care when handling these instruments to 
ensure they maintain the specied accuracy.
9.3  Calibration
Use the procedure below to check instrument 
operation and accuracy during periodic 
maintenance or troubleshooting.
9.3.1  Pressure and Static Pressure.
1)  Connect the instruments as shown in gure 9.1 
and warm up the instruments for at least ve 
minutes.
IMPORTANT
•  Do not perform the calibration procedure 
until the transmitter is at room temperature.
•  To adjust the transmitter for highest 
accuracy, make adjustments with the 
power supply voltage and load resistance 
including leadwire resistances set close to 
the conditions under which the transmitter is 
installed.
•  If the measurement range 0% point is 
0 kPa or shifted in the positive direction 
(suppressed zero), the reference pressure 
should be applied as shown in the gure.
  If the measurement range 0% point is shifted 
in the negative direction (elevated zero), the 
reference pressure should be applied using 
a vacuum pump.
2)  Apply reference pressures of 0%, 25%, 
50%, 75%, and 100% of the measurement 
range to the transmitter. Calculate the errors 
(differences between digital voltmeter readings 
and reference pressures) as the pressure is 
increased from 0% to 100% and is decreased 
from 100% to 0%, and conrm that the errors 
are within the required accuracy.










