    <5.  Installation>
5-4
IM 01C25R01-01E
5.5  Rotating Transmitter Section
The transmitter section can be rotated 
approximately 360° (180° to either direction or 
360° to one direction from the original position at 
shipment, depending on the conguration of the 
instrument.) It can be xed at any angle within 
above range.
1)  Remove the two setscrews that fasten the 
transmitter section and capsule assembly, 
using the Allen wrench. 
2)  Rotate the transmitter section slowly and stop it 
at designated position.
3)  Tighten the two setscrews to a torque of 1.5 
N·m.
IMPORTANT
Do not rotate the transmitter section more than 
the above limit.
Vertical impulse piping type
Horizontal impulse piping type
Pressure-detector section
Transmitter section
Rotate 0 to ±180° segments
Rotate 0 to ±180° segments
Transmitter section
Pressure-detector section
Conduit connection
Conduit connection
Zero-adjustment screw
Stopper
F0507.ai
Figure 5.7  Rotating Transmitter Section (Left Side 
High Pressure Type)
5.6  Changing the Direction of 
Integral Indicator 
IMPORTANT
Always turn OFF power, release pressure and 
remove a transmitter to non-hazardous area 
before disassembling and reassmbling an 
indicator.
An integral indicator can be installed in the following 
three directions.  Follow the instructions in section 
9.4 for removing and attaching the integral indicator.  
F0508.ai
Figure 5.8  Integral Indicator Direction