Manual

<9. Parameter Lists>
9-8
IM 01C25R03-01E
Relative
Index
Index Parameter Name Factory Default
Write
Mode
Explanation
38 2338 PIPE_EXPANSION_
COEF
*1
0.00001148 O/S Indicates the pipe linear expansion coefcient value.
39 2339 PIPE_REF_
TEMPERATURE
*1
20 O/S Indicates the temperature during pipe inner diameter
measurement.
40 2340 BASE_DENSITY_
FOR_VOLUME_
FLOW
*1
1.250380253 O/S Indicates the density for conversion of the volume ow unit.
41 2341 FLOW_CONFIG1
*1
O/S Indicates the compensation coefcient of the ow coefcient.
42 2342 FLOW_CONFIG2
*1
O/S Indicates the compensation coefcient of the expansion
compensation coefcient.
43 2343 FLOW_CONFIG3
*1
O/S Indicates the precision compensation coefcient.
44 2344 FLOW_CONFIG4
*1
O/S Indicates the viscosity compensation coefcient.
45 2345 CORRECTION_
VALUE
*1
Indicates the ow compensation value.
46 2346 CONFIG_SOFT_REV
*1
O/S Space Used as a record of the setup tool software version.
47 2347 CONFIG_DATE
*1
O/S Space Used as a record of the setting date.
48 2348 CONFIG_WHO
*1
O/S Space Used as a record of the setter.
49 2349 CONFIG_STATUS
*1
O/S Space Used as a memo.
50 2350 CONFIG_
VSTRING32
*1
O/S Space Used as a memo.
51 2351 CONFIG_
VSTRING16
*1
O/S Space Used as a memo.
52 2352 CONFIG_
OSTRING32
*1
O/S Space Used as a memo.
53 2353 CONFIG_OSTRING2
*1
O/S Sum value of precision ow setting data.
*1: These are the parameters which are allowed to be written only by EJXMVTool. Changes using other parameter setting tools or
handheld terminals may cause Check Sum Error (AL83).
CAUTION
Indexes 2331 to 2353 are parameters for precision volume calculation, and data calculated by the MV
setup tool must be written to these parameters. For this reason, these parameters must not be written with
data other than that from the MV setup tool.
When parameters marked by the “*1” symbol are written individually, a sum value error and an alarm
occurs.
Table 9.1 Primary Device Codes
Code Primary Device
1 Fixed Mode
2 Orice Corner Taps [ISO5167-1 1991]
3 Orice Corner Taps [ISO5167-2 2003]
4 Orice Corner Taps [ASME MFC-3M 1989]
5 Orice Flange Taps [ISO5167-1 1991]
6 Orice Flange Taps [ISO5167-2 2003]
7 Orice Flange Taps [ASME MFC-3M 1989]
8 Orice Flange Taps [AGA No.3 1992]
9 Orice D and D/2 Taps [ISO5167-1 1991]
10 Orice D and D/2 [ISO5167-2 2003]
11 Orice D and D/2 [ASME MFC-3M 1989]
12 ISA1932 nozzle [ISO5167-1 1991/ ISO5167-3 2003]
13 Long radius nozzle [ISO5167-1 1991/ ISO5167-3 2003]
14 ASME FLOW NOZZLES [ASME MFC-3M 1989]
15 Venturi nozzle [ISO5167-1 1991/ ISO5167-3 2003]
16 Classical Venturi tube “as castconvergent section [ISO5167-1 1991/ ISO5167-4 2003]
17 ASME Venturi Tubes With a rough Cast or Fabricated Convergent [ASME MFC-3M 1989]
18 Classical Venturi tube with a machined convergent section [ISO5167-1 1991/ ISO5167-4 2003]
19 ASME Venturi Tubes With a machined convergent section [ASME MFC-3M 1989]
20 Classical Venturi tube with a rough-welded sheet-iron convergent section [ISO5167-1 1991/ ISO5167-4 2003]