<Appendix 8. Advanced Diagnostics>
A8-13
IM 01C25R03-01E
The default values at the factory setting are the
values of DIAG_LIM [1] to [8] shown in Table A8. 2.
Change the threshold value to solve your problem
according to the above image
(1) Set “Stopto DIAG_MODE.
(2) Change the unsuitable value of DIAG_
LIM corresponding to the each blockage
detection.
Note:Set to “Calculation” after setting the parameters.
DIAG_LIM
# Parameter Threshold value
[1] fdpmax Threshold to detect “A Blockingby
using RATIO_FDP
[2] fdpmin Threshold to detect “B Blockingby
using RATIO_FDP
[3] fsplmax Threshold to detect “Large
Fluctuation of Low Side” by using
RATIO_FSPL
[4] fsplmin Threshold to detect “Low Side
Blockingby using RATIO_FSPL
[5] fsphmax Threshold to detect “Large
Fluctuation of High Sideby using
RATIO_FSPH
[6] fsphmin Threshold to detect “High Side
Blockingby using RATIO_FSPH
[7] blkfmax Threshold to detect “High Side
Blockingby using VALUE_BLKF
[8] blkfmin Threshold to detect “Low Side
Blockingby using VALUE_BLKF
Sampling Period
If uctuating around the threshold value, an
alarm may be often generated. When the above
phenomenon happens, the sampling time (DIAG_
PERIOD) can be changed as as to enhance the
accuracy of the blockage detection.
The longer the sampling time, better the expected
accuracy.
(1) Set “Stopto DIAG_MODE.
(2) Enter the value to DIAG_PERIOD within
the range of 20 to 65535 (s).
Note: Set to “Calculation” after setting the parameters.
Also, the accuracy can be improved by increasing
the number of DIAG_COUNT.
ILBD Range Setting
When the VALUE_DPAVG is less than the value of
the DIAG_LIM [10] or exceeds the value of DIAG_
LIM [9] shown in Table A8. 2, EJX gives an alarm of
“Outside Diagnosis Range”.
If ow/differential pressure is less than the default
threshold value of DIAG_LIM [10], pressure
uctuation is not large enough to detect the
blockage. To prevent the fault blockage detection,
the threshold value should be changed to larger
value.
(1) Set “Stopto DIAG_MODE.
(2) Enter the value to DIAG_LIM [10].
Note: Set to “Calculation” after setting the parameters.
COMP_FLG
When “Compensation is selected in COMP_FLG,
RATIO_FDP is compensated by following formula
and used as treatable monitoring value, CRATIO_
FDP.
X
REFERENCE_DPAVG
CRATIO_FDP =
VALUE_DPAVG
VALUE_FDP
REFERENCE_FDP
On the other hand, if the compensation is not
necessary, “Non-compensation” is selected in
COMP_FLG and RATIO_FDP is used as NRATIO_
FDP.
=
NRATIO_FDP
VALUE_FDP
REFERENCE_FDP
A8.2.11 Reset of Reference Value
When there are large ow change or the change of
uid under the measured process conditions, obtain
the reference value again.
If ow change by a comparison of the reference
value is ±25% or more, you need to obtain the
reference value again.