    <Appendix 6.  Link Master Functions>
A6-6
IM 01C25R03-01E
A6.5.2  Descriptions for LM Parameters
The following describes LM parameters of an EJX 
multivariable transmitter.  
NOTE:  Do not turn off the power to the EJX for 60 seconds 
after making a change to its parameter settings.  
(1)  DlmeLinkMasterCapabilitiesVariable
Bit
Position
Meaning Description Value
B3: 0x04 LAS 
Schedule in 
Non-volatile 
Memory
Whether the LAS schedule 
can (=1) or cannot (=0) be 
saved to the non-volatile 
memory
1
B2: 0x02 Last Values 
Record 
Supported
Whether to support (=1) 
or not to support (=0) 
LastValuesRecord.
0
B1: 0x01 Link Master 
Statistics 
Record 
Supported
Whether to support (=1) 
or not to support (=0) 
DlmeLinkMasterStatisticsRecord.
0
(2)  DlmeLinkMasterInfoRecord
Sub-
index
Element
Size
[bytes]
Descrip-
tion
1 MaxSchedulingOverhead 1 V(MSO)
2 DefMinTokenDelegTime 2 V(DMDT)
3 DefTokenHoldTime 2 V(DTHT)
4 TargetTokenRotTime 2 V(TTRT)
5 LinkMaintTokHoldTime 2 V(LTHT)
6 TimeDistributionPeriod 4 V(TDP)
7 MaximumInactivityToClaimLasDelay 2 V(MICD)
8
LasDatabaseStatusSpduDistributionPeriod
2 V(LDDP)
(3)  PrimaryLinkMasterFlagVariable
Explicitly declares the LAS.  Writing “true” (0xFF) 
to this parameter in a device causes that device to 
attempt to become the LAS.  However, a request 
of writing “true” to this parameter in a device is 
rejected if the value of the same parameter in any 
other device that has a smaller node address within 
the same segment is true.   
(4)  LiveListStatusArrayVariable
A 32-byte variable, in which each bit represents the 
status of whether a device on the same segment 
is live or not.  The leading bit corresponds to the 
device address 0x00, and nal bit to 0xFF.   The 
value of LiveListStatusArrayVariable in the case 
where devices having the addresses 0x10 and 
0x15 in the eldbus segment is shown below.  
0x00 00 84 00 00 00 00 00 00 00 00 00 00 00
 00 00 00 00 00 00 00 00 00 00
 00 00 00 00 00 00 00 00
Bit correspondences: 0 0 0 0 0 0 0 0 0 0 0
0 0 0 0 0 1 0 0 0 0 1 0 0...
0×00
0×10 0×15
(5)  MaxTokenHoldTimeArray
An 8×64 byte array variable, in which each set of 
2 bytes represents the delegation time (set as an 
octet time) assigned to a device.  The delegation 
time denotes a time period that is given to a device 
by means of a PT message sent from the LAS 
within each token circulation cycle.  
The leading 2 bytes correspond to the device 
address 0x00, and the nal 2 bytes to the device 
address 0xFF.  Specify the subindex to access this 
parameter.  
(6)  BootOperatFunctionalClass
Writing 1 to this parameter in a device and 
restarting the device causes the device to start as 
a basic device.  On the contrary, writing 2 to this 
parameter and restarting the device causes the 
device to start as an LM.