Instruction Manual
    <9.  Parameter Summary>
9-4
IM 01C27H01-01EN
Object
ID
Attribute 
ID
Label Description
Default
value
Handling
3.
CO
block
(continued)
4 PUB_ITEM_MAX Maximum PUB_ITEM value --- R
5 PUB_ITEM_NUM PUB_ITEM number --- R
6 PUB_ITEM Indicates the PUB_ITEM information. The following 
shows the components
1. ObjectID
2. AttributeID
3. AttributeIndex
4. Size
--- W
4.
TRANSDUCER
block
1 Tag Description Memo eld available to write anything. Transducer W
2 Auto Recovery Allows specifying the action when the cause of a 
sensor failure is removed.
OFF (=0):  AL continues to be indicated  
  even after the cause of the sensor  
  error is removed, and the BURN  
  OUT state is not released.
ON (=1):  The AL indication for sensor  
  failure disappears and normal  
  action is resumed when the cause  
  of the sensor failure is removed.
ON W
3 Model Indicates the model name of the transmitter. --- R
4 Sensor Serial 
Number
Indicates the device number of the transmitter. --- R
5 Measurement 
Rate
*3
Indicates the measurement period of process values. --- R
6 Measurement 
Mode
Measurement mode selects continuous mode or 
intermittent mode.
When the update time is set to 0.5 second
*4
 in 
intermittent mode, the mode is automatically switched 
to continuous mode.
When the update time is set to 0.5 second
*4
, 
the computation process is in continuous mode, 
regardless of the measurement mode.
When 0 is set, measurement period is set to 30 
seconds regardless of measurement setting.
intermittent 
mode
W
7 Wireless Status Indicate the communication status.
1. Indicates the initial idle status or join status.
  (idle status, join status)
2. Indicates whether Contract(Pub) is established.
  (not establishment, establishment) 
3. Indicates whether Contract(R/W) is  
established.
  (not establishment, establishment)
1. idle status
2. not 
estabilishment
3. not 
esabilishment
R
8 Display Selection Select PV Value displaying on the LCD indicator.
1. Indicates display status of PV Value of AI1
  (displayed, not displayed)
2. Indicates display status of PV Value of AI2
  (displayed, not displayed)
3. Indicates display status of PV Value of AI3
  (displayed, not displayed)
1. displayed
2. not 
displayed
3. not 
displayed
W
9 LCD Mode Select the display mode on the LCD indicator.
1. Indicates the ON mode. (off, intermittent)
2. Indicates whether continuous or not. 
  (continuous off, continuous on)
*3
3. Indicates the bar graph display.
  (bar graph not displayed, bar graph displayed)
1. OFF
2. continuous 
OFF
*3
3. bar graph 
not displayed
W
10 Special Cmd Special function parameter.
0 = initialize value at READ (None)
1 = Squawk status
2 = deep sleep status
To start from deep sleep mode, either remove and 
insert the battery pack, or use the provisioning device 
tool or the device conguration tool via infrared 
communication.
0 W