    <8.  Maintenance>
8-9
IM 01C27H01-01EN
Connect the device configuration tool and check self-diagnostics.
Contact Yokogawa service personnel.
F0808.ai
Refer to Alarm Message Summary 
in Subsection 8.5.3
Bind capillaries together as far as 
possible or provide lagging.
Provide lagging and/or heat insulation, 
or allow adequate ventilation.
Refer to 8.2 “Calibration Instruments 
Selection”. 
Adjust the output.
Does 
the ambient 
temperature significantly 
differ between the high and low 
pressure side
capillaries?
Is output adjusted correctly?
Large output error.
Does the self-
diagnostic indicate the problem
location?
Is transmitter
installed where there is 
marked variation in 
temperature?
Are excess 
capillaries secured?
YES
NO
YES
NO
NO
YES
NO
YES
NO
YES
Secure them so that they are not
moved by wind or vibration.
Were appropriate 
instruments used for 
calibration?
YES
NO