Instruction Manual
    <7.  Setting Parameters>
7-1
IM 01C27H01-01EN
7.  Setting Parameters
This transmitter can remotely handle range 
changes, Tag No. setup, monitoring of self-
diagnostic results, and zero point adjustment, etc. 
according to communication with the eld wireless 
conguration tool or the device conguration tool.
7.1  Environment for parameter 
setting
After installing the battery pack, perform 
provisioning and have the instrument join the eld 
wireless network. 
This transmitter supports the OOB (out-of-band) 
method using the infrared communication as a 
provisioning method.
Refer to Subsection 6.4 “Connecting to the Field 
Wireless Network” for details of the provisioning.
F0701.ai
Infrared port
Figure 7.1  Infrared port of the transmitter
7.2  Preparing Software
7.2.1  Softwares for the Field Wireless 
Conguration Tool and the Device 
Conguration Tool
Before using the device conguration tool, 
conrm that CF/DD and DeviceDTM for your 
transmitter (wireless EJX) are installed in the device 
conguration tool. 
CF/DD and DeviceDTM are provided by DVD-ROM 
attached to YFGW710 Field Wireless Integrated 
Gateway.
Refer to YFGW710 Field Wireless Integrated 
Gateway (IM 01W01F01-01EN) for details.
CF (Capabilities File)/DD (Device Description)
  A CF le contains information, such as the 
vendor of the eld device, its model and 
revision, available types of process data (ow 
rate, temperature, pressure, etc.), and number 
of data items. A DD le contains the information 
on parameters, such as data structures and 
attributes.
DeviceDTM
  DeviceDTM (Device Type Manager), is driver 
software for eld devices provided based on the 
FDT (Field Device Tool) technology. 
The eld wireless conguration tool or the device 
conguration tool allows conrming the device 
information. 
Refer to Subsection 8.2 “Calibration Instruments 
Selection” for the eld wireless conguration 
tool or the device conguration tool of our 
recommendation.
7.2.2  Software Download
Software download function permits to update 
wireless eld device software via ISA100.11a 
wireless communication. For details, refer to 
YFGW710 Field Wireless Integrated Gateway (IM 
01W01F01-01EN) or YFGW410 Field Wireless 
Management Station (IM 01W02D01-01EN). 
7.3  Setting Parameters
7.3.1  Parameter Usage and Selection
Before setting a parameter, please see the following 
table for a summary of how and when each 
parameter is used.
IMPORTANT
After setting and sending data with the eld 
wireless conguration tool or the device 
conguration tool, wait 30 seconds before 
turning off the transmitter. If it is turned off 
too soon, the settings will not be stored in the 
transmitter.