Manual
    <7.  Operation>
7-3
IM 01C27C01-01EN
7.3  Starting Operation
After completing the zero point adjustment, follow 
the procedure below to start operation.
1)  Conrm the operating status.
2)  After conrming the operating status, perform 
the following:
IMPORTANT
•  Close the terminal box cover and the 
amplier cover. Screw each cover in tightly 
until it will not turn further.
•  Tighten the zero-adjustment cover mounting 
screw to x the cover in position.
7.4  Connecting to the Field 
Wireless Network
■  Preparation work prior to connecting to a 
eld wireless network
This transmitter does not need to be connected 
with a physical wire. Instead of physical wiring, it 
is necessary to set the eld wireless devices to 
communicate with before installing the transmitter. 
This procedure is called a provisioning.
This transmitter supports provisioning via infrared 
communication using a provisioning device and 
can be securely connected to a network. If the 
provisioning information is not set, the transmitter 
cannot be connected to the eld wireless network.
Provisioning:
Provisioning is work to set the security and 
network information to enable the transmitter 
to be connected to the eld wireless network. 
This transmitter supports a provisioning method 
using infrared communication.
For details on provisioning using a provisioning 
device, and procedure for connecting the 
transmitter to the Field Wireless Network, refer 
to the User’s Manual, FieldMate Versatile Device 
Management Wizard (IM 01R01A01-01E), 
YFGW710 Field Wireless Integrated Gateway (IM 
01W01F01-01EN) and YFGW410 Field Wireless 
Management Station (IM 01W02D01-01EN).
F0704.ai
within 30cm
Figure 7.4  Provisioning Example
■  Provisioning work
This subsection describes provisioning work using 
FieldMate as the provisioning device.
Provisioning work performs provisioning for each 
eld wireless device using FieldMate and an 
infrared adapter. If use Yokogawa-recommended 
infrared device for provisioning, set distance 
between the glass window surface to the infrared 
device within 30cm.
For details of Yokogawa-recommended infrared 
device, refer to subsection 9.2 “Calibration 
Instruments Selection”.
Perform the following provisioning tasks. 
•  Setting provisioning information
•  Creating a provisioning information le
1) Setting provisioning information
  Set the device tag and network ID using a 
FieldMate provisioning function. The device 
tag, network ID, and join key are set in the eld 
wireless device. It is not necessary to input 
a join key because FieldMate automatically 
generates it.
Setting device tag
The device tag is used for the user to 
recognize the eld wireless device.
Setting network ID
This is the network ID for the eld wireless 
network to which the eld wireless device is 
connected. Set a value from 2 to 65535. 
The eld wireless device is connected to the 
eld wireless network corresponding to the 
network ID set by provisioning work.