User guide
    <Appendix 1.  Safety Instrumented Systems Installation>
A1-1
IM 01C25T03-01E
Appendix 1.  Safety Instrumented Systems 
Installation
WARNING
The contents of this appendix are cited from 
exida.com safety manual on the transmitters 
specically observed for the safety transmitter 
purpose. When using the transmitter for Safety 
Instrumented Systems (SIS) application, the 
instructions and procedures in this section must 
be strictly followed in order to preserve the 
transmitter for that safety level.
A1.1  Scope and Purpose
This section provides an overview of the user 
responsibilities for installation and operation of the 
transmitter in order to maintain the designed 
safety level for Safety Instrumented Systems (SIS) 
applications. Items that will be addressed are proof 
testing, repair and replacement of the transmitter, 
reliability data, lifetime, environmental and 
application limits, and parameter settings.
A1.2  Using the Transmitter for 
an SIS Application
A1.2.1  Safety Accuracy
The transmitter has a specied safety accuracy 
of 2%. This means that the internal component 
failures are listed in the device failure rate if they will 
cause an error of 2% or greater.
A1.2.2  Diagnostic Response Time
The transmitter will report an internal failure within 5 
seconds of the fault occurrence.
A1.2.3  Setup
During installation the transmitter must be setup 
with engineering units parameters. This is typically 
done with a handheld terminal. These parameters 
must be veried during the installation to insure 
that the correct parameters are in the transmitter. 
Engineering range parameters can be veried by 
reading these parameters from the optional local 
display or by checking actual calibration of the 
transmitter.
The calibration of the transmitter must be performed 
after parameters are set.
A1.2.4  Required Parameter Settings
The following parameters need to be set in order to 
maintain the designed safety integrity.
Table A1.1  Required Parameter Settings
Item Description
Burnout 
direction switch
To specify if the output should go 21.6 
mA or higher or 3.6 mA or lower upon 
detection of an internal failure.
Write protection 
switch
The write function should be 
disabled.
A1.2.5  Proof Testing
The objective of proof testing is to detect failures 
within the transmitter that are not detected by the 
diagnostics of the transmitter. Of main concern 
are undetected failures that prevent the safety 
instrumented function from performing its intended 
function. See table A1.2 for proof testing method.
The frequency of the proof tests (or the proof 
test interval) is to be determined in the reliability 
calculations for the safety instrumented functions 
for which the transmitter is applied. The actual 
proof tests must be performed more frequently or 
as frequently as specied in the calculation in order 
to maintain required safety integrity of the safety 
instrumented function.
The following tests need to be specically executed 
when a proof test is performed. The results of 
the proof test need to be documented and this 
documentation should be part of a plant safety 
management system. Failures that are detected 
should be reported to Yokogawa.
The personnel performing the proof test of the 
transmitter should be trained in SIS operations 
including bypass procedures, transmitter 
maintenance, and company management of 
change procedures.










