User guide
    <3.  Operation>
3-13
IM 01C25T03-01E
(12) Output Status Setup when a Hardware 
Error Occurs 
(D26: ERROR OUT)
This parameter allows the setting of the output 
status when a hardware error occurs. The following 
selections are available.
(a)  BURNOUT DIR; Outputs the corresponding 
values of 110% or –5% of output signals 
according to the setting by burnout direction 
switch (BO) on the CPU board.
(b)  HOLD; Outputs the last value held before the 
error occurred.
Note:  A hardware error means CAP.ERR of AL.01 or AMP.ERR 
of AL.02 which are shown in table 4.1 Alarm Message 
Summary.
• Example: Set the output status to HOLD when 
   a hardware error occurs.
ESC
SET
 D26:ERROR OUT
 BURNOUT DIR
 < BURNOUT DIR >
 < HOLD >
Use the or key 
to select HOLD.
Press the key twice to 
enter the setting.
F0331.ai
Then press the (OK) key.
(13) Bi-directional Flow Measurement 
Setup
(E30: BI DIRE MODE)
(a)  This parameter enables selection of 50% output 
at an input of 0 kPa.
  Procedure is shown in the gure below.
(b)  Combining this with C40: OUTPUT MODE 
provides a square root output computed 
independently for 0% to 50% output and for 
50% to 100% output.
• Example: If measurement range is 0 to 10 kPa 
   (LRV=0 kPa, URV=10 kPa)
The measurement range changes to –10 to 0 to 10 kPa 
(output 0% to 50% to 100).
Note that C21: PRES LRV and C22: PRES URV 
are not changed.
ESC
SET
 E30:BI DIRE MODE
 OFF
 < OFF >
 < ON >
Use the or key 
to select ON.
Press the key twice to 
enter the setting.
F0332.ai
Then press the (OK) key.
20 mA (100% display)
4 mA (–100% display)
● Output mode “LINEAR”
LRV HRV
F0333.ai
20 mA (100% display)
Low Cut
4 mA (–100% display)
● Output mode “SQUARE ROOT”
LRV HRV
(14) Range Change while Applying Actual 
Inputs 
(H10: AUTO P LRV, H11: AUTO P URV)
This feature allows the lower and upper range 
values to be set up automatically with the actual 
input applied. If the lower and upper range values 
are set, C21: PRES LRV and C22: PRES URV are 
changed at the same time.
Follow the procedure in the gure below. 
The measurement span is determined by the upper 
and lower range values. Changing the lower range 
value results in the upper range value changing 
automatically, keeping the span constant.
• Example 1: When changing the lower range 
   value to 0.5 kPa for the present 
setting of 0 to 30 kPa, take the 
following action with input pressure 
of 0.5 kPa applied.
FEED NO OK
SET
 H10:AUTO P LRV
 0.5000 kPa
DATA DIAG PRNT ESC
PARAM
 H10:AUTO P LRV
 0.5000 kPa
 H11:AUTO P URV
 30.500 kPa
 H20:AUTO SP LRV
 0.0 MPa
ESC
SET
 H10:AUTO P LRV
 0 kPa
 + 0
F0334.ai
The upper range value is changed 
keeping the span constant.
Parameters C21 and C22 are 
changed at the same time.
Press the key twice. 
The lower range value is changed 
to 0.5 kPa.
Press the (OK) key.