Manual
    <4.  Diagnostics>
4-29
IM 01C25T01-06EN
Integral 
Indicator
HART conguration 
 tool display
Device Status
Extended 
Device Status
Value and Status (Data Quality and Limit Status)
Differential 
Pressure 
(DP)
Static 
Pressure 
(SP)
Temperature 
(T)
% range Loop current
AL.35 
P. HI
P high alarm *
1
Pressure High 
Alarm
 *
2
--- Device 
Variable Alert
(0x02)
Value: Measured value 
Status: Good and Not Limited
AL.36 
P. LO
P low alarm *
1
Pressure Low 
Alarm 
*
2
AL.37 
SP. HI
SP high alarm *
1
Static Pressure 
High Alarm *
2
AL.38 
SP. LO
SP low alarm *
1
Static Pressure 
Low Alarm *
2
AL.50 
P. LRV
Illegal P LRV *
1
Illegal Pressure 
LRV *
2
Value: Measured value 
Status: Good and Not Limited
Value: Hold value 
Status: Bad and Constant
AL.51 
P. URV
Illegal P URV *
1
Illegal Pressure 
URV *
2
AL.52 
P. SPN
Illegal P SPAN *
1
Illegal Pressure 
SPAN *
2
AL.53 
P. ADJ
P SPAN trim err *
1
Pressure SPAN 
Trim Error *
2
Value: 
Measured 
value 
Status:  
Poor 
Accuracy and 
Not Limited
Value:  
Measured value 
Status:  
Good and Not Limited
Value: Measured value 
Status: Poor Accuracy and 
Not Limited
P ZERO trim 
err *
1
Pressure ZERO 
Trim Error *
2
AL.54 
SP. RNG
Illegal SP LRV *
1
Illegal Static 
Pressure LRV *
2
Value: Measured value 
Status: Good and Not Limited
Value: Measured value 
Status: Good and Not 
Limited
Illegal SP URV *
1
Illegal Static 
Pressure URV *
2
Illegal SP SPAN 
*
1
Illegal Static 
Pressure SPAN *
2
AL.55 
SP. ADJ
SP SPAN trim 
err *
1
Static Pressure 
SPAN Trim Error 
*
2
Value:  
Measured 
value 
Status:  
Good and Not  
Limited
Value:  
Measured 
value 
Status:  
Poor 
Accuracy and 
Not Limited
Value:  
Measured 
value 
Status:  
Good and Not 
Limited
Value: Measured value 
Status: Good and Not 
Limited
SP ZERO trim 
err *
1
Static Pressure 
ZERO Trim Error 
*
2
AL.60
SC.CFG
SC cong error Value: Measured value
Status: Good and Not Limited
AL.79 
OV.DISP
--- --- --- Value: Measured value 
Status: Good and Not Limited
*1:  DD and DTM (excluding EJX HART 5 DTM based on FDT1.2)
*2:  EJX HART 5 DTM based on FDT1.2
*3:  Depend on the setting of hardware switch
*4:  Depend on the direction of range over (high or low)