Manual
    <2. Connection>
2-2
IM 01C25T01-06EN
(a)  Conrmation on the name plate
  The last numerical number engraved after 
Serial number and year of production shows 
HART protocol revision number at the shipment 
which is shown in Figure 2.1.
F0201.ai
: Refer to USER'S MANUAL.
Made in Japan
TOKYO 180-8750 JAPAN
MODEL
SUFFIX
SUPPLY
OUTPUT
MWP
mA DC
V DC
STYLE
CAL
RNG
NO.
XXX - - - XX 
Revision No.
Output Signal 
Code
Revision No.
HART Protocol 
Revision
J
5 HART 5
7 HART 7
- HART 7
E - HART 5
Figure 2.1   Hart Protocol Revision Number on 
Name Plate
(b)  Conrmation on integral indicator  
(A case of integral indicator code D is specied)
  Refer to section 2.1.
(c)  Conrmation by using HART conguration tool
1) Connect the conguration tool to the 
transmitter.
2) Conrm numerical number displayed on 
“Universal rev” column.
  • Procedure to call up the display
DD and DTM 
(excluding EJX_
HART 5[1.2])
[Root Menu (refer to 3.1.1) ] 
→ Review → Universal rev
EJX_HART 5[1.2] 
DTM
Conguration → HART 
→Universal rev
IMPORTANT
Protocol revision supported by HART 
conguration tool must be the same or higher 
than that of the transmitter.
Protcol revision 
supported by HART 
conguration tool
5 7
Protocol revision of 
the transmitter
5
 
7 ×
  : Communication OK
×: Communication NG
NOTE
HART 7 communication is supported by 
FieldMate R2.02 or later.
NOTE
When the output signal code of the transmitter 
is “-J”, HART protocol revision can be changed.  
Refer to subsection 3.3.18 about the procedure 
of the revision change of HART 5 and HART 7.
2.3  Device Description (DD) on 
a Conguration Tool and 
Transmitter Device Revision
Before using a HART conguration tool, conrm 
that the DD for the transmitter is installed in the 
conguration tool.
About the DD, use the device type, device revision 
and DD Revision shown in the Table 2.1. 
Table 2.1  HART Protocol Revision, Device 
Revision and DD Revision
HART 
Protocol 
Revision 
(*1)
DPharp Pressure Transmitter
Model
Device 
Type
Device 
Revision
DD
Revision
5
EJX series
EJX
(0x51)
3
3
or later
EJA series
EJA-NEXT
(0x5C)
1
2
or later
7
EJX series
EJX
(0x3751)
10
2
or later
EJA series
EJA-NEXT
(0x375C)
10
1
or later
*1:  When the output signal code is “–E”, only “5” is available.
The device revision of the transmitter and DD can 
be conrmed as shown below.
If the correct DD is not installed in the conguration 
tool, download it from the ofcial web site of HART 
Communication Foundation.
(1)  Conrmation of device revision for the 
transmitter
●  Conrmation on integral indicator  
(A case of integral indicator code D is 
specied)
  Refer to the section 2.1