Manual
    <2.  Connection>
2-1
IM 01C25T01-06EN
2.  Connection
2.1   Integral Indicator Display 
When Powering On
For models with the integral indicator code “D”, the 
display shows all segments in the LCD and then 
changes to the displays shown below sequentially.
All segments display 
Model name (3 s)
Communication Protocol (3 s) 
Device Revision (3 s) 
F0200.ai
Either “5” or “7” is displayed on the “Communication 
Protocol” display as HART protocol revision 
followed by device revision number on the “Device 
Revision” display.
NOTE
This function is available for software revision 
2.02 or later.
Software revision can be checked by the 
following procedure.
DD and DTM 
(excluding EJX_
HART 5[1.2])
[Root Menu](Refer to subsection 
3.1.1) → Review → Software rev 
EJX_HART 5[1.2] 
DTM
Conguration → Device 
information1 → Software rev
NOTE
In this User’s Manual, HART protocol revision 
5 and 7 are described as HART 5 and HART 7 
respectively.
NOTE
LCD display can be set to “All segments display” 
only.
• Procedure to call up the display 
DD (HART 5/7)
DTM (HART 7 
/EJA:HART 5)
[Root Menu] (Refer to subsection 
3.1.1) → Detailed setup → Display 
condition → Chg power on info 
DTM  
(EJX:HART 5)
Conguration → Local Display 
→ Chg power on info
ON
Show all segments display, Model 
name, Communication Protocol, and 
Device Revision when powering on.
OFF
Show all segments display when 
powering on.
2.2    HART Protocol Revision
For the models with the output signal code “-J”, 
HART protocol revision 5 or 7 is selectable. The 
protocol revision is set as specied in the order.  
The typical function which is available by HART 
protocol revision 7 is listed as follows. Refer to 
HART 7 description in this document or 
HART 7
 mark 
for detail.
•  Long Tag Supporting Up to 32 Characters
  Long tag secures a better asset management 
with abundant digits in its software.
•  Enhanced Burst Mode and Event Notication 
  Advanced burst mode includes the variety 
of transmission setting by specifying burst 
variables, update period, and message trigger 
mode, and event notication function gives 
you alert signal based on the status change in 
preset values and self-diagnosis.
•  Squawk
  Identifying the transmitter by displaying the 
particular pattern on LCD
•  Multidrop Communication
  Up to 63 transmitters can be connected. An 
analog signal output available for one device in 
a loop.
How to conrm protocol revision is shown below.
There are three ways to conrm the protocol 
revision set to the transmitter.