Manual

<9. Parameter Lists>
9-7
IM 01C25T04-01EN
Relative
Index
Index Parameter
Write
Mode
Valid
Range
Initial Value Explanation
13 29 UNIT_SEL1 AUTO 0-1 Space Selection of unit to be displayed. The unit of the
parameter which is selected at PARAMETER SEL1
will be displayed when "Auto" is selected; user-
specied unit at DISPLAY UNIT1 will be displayed
when "Custom" is selected.
14 30 DISPLAY_UNIT1 AUTO Space User specied unit to be displayed on display1,
which will be available when "Custom" is selected at
UNIT SEL1.
15 31 EXP_MODE1 AUTO 0-3 0: Engineering
Unit
Selection of the displayed value in exponent such as
x1, x10, x100 and x1000.
16 32 BLOCK_TAG2 STB Block tag which includes a parameter to be
displayed on display2
17 33 PARAMETER_SEL2 AUTO 0x0512:
Primary Value
Selection of a parameter to be displayed on
display2. Select a parameter from Table
18 34 DISPLAY_TAG2 AUTO Space Name of block tag to be displayed on display2; up to
six alphanumeric plus a slash[/] and a period [.]
19 35 UNIT_SEL2 AUTO 0-1 0: Auto Selection of unit to be displayed. The unit of the
parameter which is selected at PARAMETER SEL2
will be displayed when "Auto" is selected; user-
specied unit at DISPLAY UNIT2 will be displayed
when "Custom" is selected.
20 36 DISPLAY_UNIT2 AUTO Space User specied unit to be displayed on display2,
which will be available when "Custom" is selected at
UNIT SEL2.
21 37 EXP_MODE2 AUTO 0-3 0: Engineering
Unit
Selection of the displayed value in exponent such as
x1, x10, x100 and x1000.
22 38 BLOCK_TAG3 STB Block tag which includes a parameter to be
displayed on display3
23 39 PARAMETER_SEL3 AUTO 0x0542: Static
Pressure(H)
Selection of a parameter to be displayed on
display3. Select a parameter from Table
24 40 DISPLAY_TAG3 AUTO Space Name of block tag to be displayed on display3; up to
six alphanumeric plus a slash[/] and a period [.]
25 41 UNIT_SEL3 AUTO 0-1 0: Auto Selection of unit to be displayed. The unit of the
parameter which is selected at PARAMETER SEL3
will be displayed when "Auto" is selected; user-
specied unit at DISPLAY UNIT3 will be displayed
when "Custom" is selected.
26 42 DISPLAY_UNIT3 AUTO Space User specied unit to be displayed on display3,
which will be available when "Custom" is selected at
UNIT SEL3.
27 43 EXP_MODE3 AUTO 0-3 0: Engineering
Unit
Selection of the displayed value in exponent such as
x1, x10, x100 and x1000.
28 44 BLOCK_TAG4 STB Block tag which includes a parameter to be
displayed on display4
29 45 PARAMETER_SEL4 AUTO 0x051B:
Capsule
Temperature
Selection of a parameter to be displayed on
display4. Select a parameter from Table
30 46 DISPLAY_TAG4 AUTO Space Name of block tag to be displayed on display4; up to
six alphanumeric plus a slash[/] and a period [.]
31 47 UNIT_SEL4 AUTO 0-1 0: Auto Selection of unit to be displayed. The unit of the
parameter which is selected at PARAMETER SEL4
will be displayed when "Auto" is selected; user-
specied unit at DISPLAY UNIT4 will be displayed
when "Custom" is selected.
32 48 DISPLAY_UNIT4 AUTO Space User specied unit to be displayed on display4,
which will be available when "Custom" is selected at
UNIT SEL4.
33 49 EXP_MODE4 AUTO 0-3 0: Engineering
Unit
Selection of the displayed value in exponent such as
x1, x10, x100 and x1000.
34 50 BAR_GRAPH_
SELECT
AUTO 0-1 1: Enable Selection of bar graph indicator
0: Disable
1: Enable
35 51 DISPLAY_CYCLE AUTO 0: 2.8 s Duration of display cycle. (Time unit: 1=400 ms)
36 52 POWER_ON_INFO AUTO 0-1 1: ON Display when powering on.
37 53 LCD_TEST Not used for the transmitter
38 54 SW_TEST Not used for the transmitter