Manual

<5. Parameter Summary>
5-4
IM 01C25T03-01E
No. Parameter name Item
*1
R/W
Content Default value
Applicable
model
Upload
data
F P L
J ADJUST Adjusting data
J09 ADJ UNIT Pressure adjusting unit select W % or PRES UNIT PRES UNIT
J10 ADJ PRES Adjustment reference pressure R Unit specied in J09
J11 P ZERO ADJ Automatic zero adjustment W -32000 to 32000, unit
specied in J09
0.00000 kPa
J12 P SPAN ADJ Automatic span adjustment W -32000 to 32000, unit
specied in J09
100.000 kPa
J15 P ZERO DEV Manual zero adjustment W -32000 to 32000, unit
specied in J09
100.000 kPa
J16 P SPAN DEV Manual span adjustment W -32000 to 32000, unit
specied in J09
100.000 kPa
J20 ADJ SP Adjustment reference pressure
of SP
R Unit specied in J09
J21 SP ZERO ADJ Automatic SP zero adjustment W -32000 to 32000, unit
specied in J09
0.00000 MPa
J22 SP SPAN ADJ Automatic SP span
adjustment*
2
W -32000 to 32000, unit
specied in J09
J25 SP ZERO DEV Manual SP zero adjustment W -32000 to 32000, unit
specied in J09
0.00000 MPa
J26 SP SPAN DEV Manual SP span adjustment W -32000 to 32000, unit
specied in J09
0.00000 MPa
J40 OUTPUT 4mA 4 mA adjustment W -10.000 to 10.000% 0.000%
J41 OUTPUT 20mA 20 mA adjustment W -10.000 to 10.000% 0.000%
J45 AMP TEMP Amplier temperature R Unit specied D40 0 degC
J50 ADJ WHO Adjustment information W 8 alphanumeric characters
J51 ADJ DATE Adjustment information W 16 alphanumeric characters
J52 ADJ LOC Adjustment information W 8 alphanumeric characters
J53 ADJ DESC Adjustment information W 16 alphanumeric characters
J55 EXT ZERO ADJ External zeroing permission W INHIBIT or ENABLE ENABLE
J56 CLEAR ADJ Clear adjustment W ---, PRES, SP, 4-20mA, or
ALL
---
J60 SELF CHECK Self-diagnostics R See A60
K TEST Test parameters
K10 OUTPUT X % Test output % setting W Within a range between D20
and D21
0.00%
K40 DO TEST Test contact output W OFF or ON OFF
K45 TEST TIME OUTPUT X %” and “DO
TEST” duration time selection
W 10 min, 30 min, 60 min,
3 hour, 6 hour, 12 hour
10 min
K50 TEST KEY1 Special maintenance
parameter
W
K51 TEST KEY2 Special maintenance
parameter
W
K52 TEST KEY3 Special maintenance
parameter
W
K53 TEST KEY4 Special maintenance
parameter
W
K60 SELF CHECK Self-diagnostics R See A60
M DEVICE INFO Device information
M10 SERIAL NO. Serial number R
M11 MFTR. DATE Manufactured date R
M12 EXTRA NO. Customization number R
M15 SOFT REV Software revision R
M16 BRAIN REV BRAIN protocol revision R
M17 MEMO1 Memo W 16 alphanumeric characters
M18 MEMO2 Memo W 16 alphanumeric characters
M19 MEMO3 Memo W 16 alphanumeric characters
M20 ISOL MATL Capsule material W
M21 FILL FLUID Fill uid W
*1: R/W: R = Read only, W = Read & Write
*2: The default value shows MWP (Maximum working pressure) of the capsule.
Since the working pressure limit varies according to the Model, refer to the General Specications section in each user’s manual.