User guide
IM 1C22H1-01E
7-10
7. BRAIN TERMINAL BT200 OPERATION
(6) Integral Indicator Scale Setup
The following 5 displays are available for integral
indicators.
T0708.EPS
D20: DISP SELECT
and Display
NORMAL %
USER SET
USER & %
INP PRES
PRES & %
Indicates –5 to 110% range 
depending on the Measurement 
range (C21, C22).
Indicates values depending on the 
Engineering range (D22, D23).
(Note 1) 
Units set using Engineering unit 
(D21) are not indicated.
Indicates user set and % alternately 
in 3 second intervals.
Indicates input pressure. Indication 
limits –19999 to 19999.
Indicates input pressure and % 
alternately in 3 second intervals.
A10:OUTPUT (%)
45.6 %
Description
and Related parameters
A11:ENGR.OUTPUT
20.0 M
A10:OUTPUT (%)
45.6 %
A11:ENGR. OUTPUT
20.0 M
A40:INPUT
456 kPa
A10:OUTPUT (%)
45.6 %
A40:INPUT
456 kPa
(Note 1) Scale range can be specified with range limit specifications up 
to 5 digits (excluding any decimal point) for low or high range 
limits within the range of –19999 to 19999.
See (a.) through (c.) for each setting procedure.
% indication and 
input pressure 
indication
Transmitter is set 
for “% display” when 
shipped.
D20: DISP SELECT
 NORMAL %
 INP PRES
 PRES & %
User-set engineering 
unit display
D20: DISP SELECT
 USER SET
 USER & %
D21: DISP UNIT
D22: DISP LRV
D23: DISP HRV
Set for user-set 
engineering unit display.
Set a unit to be 
displayed on the BT200.
Set a numeric value for 
engineering unit for 4 mA 
output (LRV).
Set a numeric value for 
engineering unit for 20 mA 
output (HRV).
F0718.EPS
a. Display Selection (D20: DISP SELECT)
Follow the instructions given to the below to change
the range of integral indication scales.
When USER SET is selected, the user set values of
integral indication and A11: ENGR. OUTPUT
parameter are indicated.
• Example: Set the integral indicator scale to 
 engineering units display.
The “%” disappears from the 
integral indicator display.
ESC
SET
 D20:DISP SELECT
 USER SET
FEED NO OK
SET
 D20:DISP SELECT
 NORMAL %
 <NORMAL %>
 <USER SET>
 <USER & %>
 <INP PRES>
Use the or key 
to select “USER SET.”
Press the key twice to 
enter the setting.
Press the (OK) key.
F0720.EPS