Manual
IM 01C22K01-01E
10-4
10. GENERAL SPECIFICATIONS
10.3 Optional Specifications
For FOUNDATION Fieldbus explosion protected type, see IM 01C22T02-01E.
For PROFIBUS PA explosion protected type, see IM 01C22T03-00E.
Item
Description
Code
Factory Mutual (FM)
CENELEC ATEX (KEMA) Intrinsically safe Approval 
*2 
  Certificate: KEMA 02ATEX1030X
  II 1G EEx ia IIC T4, Amb. Temp.: –40 to 60C (–40 to 140F)
  Ui=30 V, Ii=165 mA, Pi=0.9 W, Ci=22.5 nF, Li=730 H
KS2
FM Intrinsically safe Approval 
*1 
  Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1, 
  Groups E, F & G and Class III, Division 1 Hazardous Locations.
  Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division. 2, 
  Groups E, F & G, and Class III, Division 1 Hazardous Locations.
  Enclosure: “NEMA 4X”, Temp. Class: T4, Amb. Temp.: –40 to 60C (–40 to 140F)
  Intrinsically Safe Apparatus Parameters
    [Groups A, B, C, D, E, F and G]
      Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 H
    [Groups C, D, E, F and G] 
      Vmax=30 V, Imax=225 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 H
FS1
CENELEC ATEX (KEMA) Flameproof Approval 
*2 
  Certificate: KEMA 02ATEX2148
  II 2G EExd IIC T4, T5, T6 
  Amb. Temp.: T5; –40 to 80C ( –40 to 176F), T4 and T6; –40 to 75C ( –40 to 167F)
  Max. process Temp.: T4; 120C (248F), T5; 100C (212F), T6; 85C (185F)
KF2
FM Explosionproof Approval 
*1 
  Explosionproof for Class I, Division 1, Groups B, C and D
  Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
  Hazardous (classified) locations, indoors and outdoors (NEMA 4X)
  Division 2, ‘SEALS NOT REQUIRED’, 
  Temp. Class: T6 
    Amb. Temp.: –40 to 60C (–40 to 140F)
FF1
FU1
Combined FF1 and FS1 
*1 
Combined KF2, KS2 and Type n 
*2 
  Type n
  II 3G Ex nL IIC T4, Amb. Temp.: –40 to 60°C (–40 to 140°F)
  Ui=30 V DC, Ci=22.5 nF, Li=730 µH
  Dust
  II 1D maximum surface temperature T65°C (149F) {Tamb.: 40°C (104F)}, 
  T85°C (185F) {Tamb.: 60°C (140F)}, T105°C (221F) {Tamb.: 80°C (176F)}
KU2
CENELEC ATEX
T1005-1.EPS
*1: Applicable for Electrical connection code 2 and 7 (1/2 NPT female).
*2: Applicable for Electrical connection code 2, 4, 7 and 9 (1/2 NPT and M20 female).