Manual
IM 01C22K01-01E
8-15
8. BRAIN TERMINAL BT200 OPERATION
• Example: For the range of 0 to 30 kPa.
  A40: INPUT = 30.15 kPa
  J15: SPAN ADJ = 0.15 % 
Suppose that a standard pressure of 30 kPa is applied and 
the value of the parameter of A40: INPUT is 30.15 kPa.
Firstly, obtain the slope error for the span as follows; 
Add 0.5% to 0.15% of the current value to calculate the 
accumulated span adjustment value.
0.15  (0.50) = 0.35 
F0847.EPS
FEED NO OK
SET
 J15:SPAN ADJ
 -0.35 %
ESCCLRDEL
SET
 J15:SPAN ADJ
 0.15 %
 - 0.35
Set 0.35.
Press key twice. 
Press the (OK) key.
30.0030.15
30.00
 100 = 0.5 (%)
Span adjustment value (%) =
Note: Enter 0.00 to J15: SPAN ADJ to reset the 
span adjustment to the initial value at the 
shipment.
P
1
 A40
P
1
100
(16) Test Output Setup (K10: OUTPUT X%)
This feature can be used to output a fixed current from
3.2 mA (–5%) to 21.6 mA (110%) for loop checks.
• Example: Output 12 mA (50%) fixed current.
ESC
SET
 K10:OUTPUT X %
 50.0 % ACTIVE
FEED NO OK
SET
 K10:OUTPUT X %
 0.0 %
 + 050.0
Set “50.0%.”
Press the key twice to 
output a fixed current at 50%.
“Active” is displayed while this is 
being executed.
Press the (OK) key to cancel 
the fixed current output.
F0832.EPS
Note: When optional code /F1 is specified, output range is from
3.6 mA(–2.5%) to 21.6 mA(110%).
IMPORTANT
1. Test output is held for approximately 10
minutes, and then released automatically after
the time has elapsed. Even if the BT200 power
supply is turned off or the communication cable
is disconnected during test output, it is held for
approximately 10 minutes.
2. Press the 
 (OK) key to release test output
immediately.
(17) User Memo Fields (M: MEMO)
This feature provides 5 user memo fields, each holding
up to 8 alphanumeric characters. Up to 5 items such as
inspection date, inspector, and other information can be
saved in these fields.
• Example: Save an inspection date of January 
 30, 1995.
DATA DIAG PRNT ESC
SET
 M10:MEMO 1
 95.1.30_
ESC
PARAM
 M10:MEMO 1
 M20:MEMO 2
 M30:MEMO 3
Set “95.1.30” in the order of year, 
month, and day.
Press the key twice to 
enter the setting.
F0833.EPS