Operation Manual

<9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>
9-11
IM 01F06A00-01EN
Simulation Setting and Correlation of Output Value
<Case A>: Without option code /MV
Simulation
Setting value
Output value
Flow rate
Total ow rate
Process temperature
Process density
Contact output
PV % Range
Loop Current
Flow rate Yes Yes No No Yes Yes Yes
Total ow rate NoYesNoNoNoNoNo
Process
temperature
NoNoYesNoNoNoNo
Process density No No No Yes No No No
PV % Range No No No No No Yes Yes
Loop Current No No No No No No Yes
Yes : Simulation value or calculation result of Simulation.
No : Actual process value or parameter setting value.
<Case B>: With option code /MV
<Case B-1>: Function == “Monitor only”
<Case B-2>: Function == “Saturated Steam” or “Superheat
Steam” or “LIQUID:Mass”
<Case B-3>: Function == “Gas:STD/Normal”
<Case B-4>: Function == “Not use”
Simulation
Setting value
Output value
Flow rate Total ow rate
Process temperature
Process density Contact output
Flow rate Yes Yes No No Yes
Total ow rate NoYesNoNoNo
Process
temperature
<Case B-1, 4>
No
<Case B-1, 4>
No
Yes <Case B-1, 2>
Yes
<Case B-1, 4>
No
<Case B-2, 3>
Yes
<Case B-2, 3>
Yes
<Case B-3, 4>
No
<Case B-2, 3>
Yes
Process density
<Case B-1, 3, 4>
No
<Case B-1, 3, 4>
No
No Yes <Case B-1, 3, 4>
No
<Case B-2>
Yes
<Case B-2>
Yes
<Case B-2>
Yes
PV % Range No No No No No
Loop Current No No No No No
Simulation
Setting value
Output value
PV= Instaneous ow rate PV= Process temperature
PV % Range Loop Current PV % Range Loop Current
Flow rate
Yes Yes <Case B-1, 2, 3>
No
<Case B-1, 2, 3>
No
<Case B-4>
Not available
<Case B-4>
Not available
Total ow rate
No No <Case B-1, 2, 3>
No
<Case B-1, 2, 3>
No
<Case B-4>
Not available
<Case B-4>
Not available
Process
temperature
<Case B-1, 4>
No
<Case B-1, 4>
No
<Case B-1, 2, 3>
Yes
<Case B-1, 2, 3>
Yes
<Case B-2, 3>
Yes
<Case B-2, 3>
Yes
<Case B-4>
Not available
<Case B-4>
Not available
Process density
<Case B-1, 3, 4>
No
<Case B-1, 3, 4>
No
<Case B-1, 2, 3>
No
<Case B-1, 2, 3>
No
<Case B-2>
Yes
<Case B-2>
Yes
<Case B-4>
Not available
<Case B-4>
Not available
PV % Range
Yes Yes <Case B-1, 2, 3>
Yes
<Case B-1, 2, 3>
Yes
<Case B-4>
Not available
<Case B-4>
Not available
Loop Current
No Yes <Case B-1, 2, 3>
No
<Case B-1, 2, 3>
Yes
<Case B-4>
Not available
<Case B-4>
Not available