Operation Manual
    <7. OPERATION FOR THE BRAIN TERMINAL (BT200)>
7-3
IM 01F06A00-01EN
7.3  Setting Parameters using BT200
This section describes the setting method using a BRAIN TERMINAL (BT200). For details on the method, refer 
to Section 6.3 “Parameters List” and Section 6.4 “Parameters Description”.
(1)  Setting Flow Span
Example : Change fl ow span 100m
3
/h to 150m
3
/h
F0704.ai
Move the inverse video bar 
to “B10 : FLOW SPAN”.
The data setup screen is 
displayed.
If the security screen is 
displayed, enter the security 
code.
Enter “150” using the 
alphanumeric keys.
Pressing the enter key, 
inverse video bar is flashed.
or
The data setting can be 
completed.
If data setting is missing, 
press “F3” key and reset the 
data.
Setting is completed.
The screen returns to the 
preceding screen when the 
“F4” (ESC) key is pressed.
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:AUX. SETUP
E:METER SETUP
H:ADJUST
HOME SET ADJ ESC
PARAM
B10:FLOW SPAN
 100 m3/h
B15:DAMPING
 4 sec
B20:CONTACT OUT
 SCALED PULSE
DATA DIAG PRNT ESC
PARAM
B10:FLOW SPAN
 100 m3/h
B15:DAMPING
 4 sec
B20:CONTACT OUT
 SCALED PULSE
DATA DIAG PRNT ESC
SET
B10:FLOW SPAN
 100 m3/h
 0
 DEL CLR ESC
SET
B10:FLOW SPAN
 100.00 m3/h
 150
 DEL CLR ESC
SET
B10:FLOW SPAN
 100.00 m3/h
 150
print off
F2:printer on
FEED POFF NO
SET
B10:FLOW SPAN
 150.00 m3/h
FEED NO OK
Enter menu item, and press 
“ ” key.
Move the inverse video bar 
to “B : EASY SETUP”.
IMPORTANT
After setting a parameter, keep the power on for 
at least 30 seconds.
If the power of fl owmeter is turned off, a 
parameter setting is released.