Operation Manual
<6. PARAMETERS>
6-3
IM 01F06A00-01EN
(3) Item C : BASIC SETUP
These items are for the basic parameters with setting before shipment.
The parameters, C20 to C50, are not indicated when option code “/MV” is selected and parameter item is
selected in F10 except “Monitor only” or “Not use”.
A value in “( )” is the data corresponding to the indicator.
Item Name R / W Data Range Unit
Decimal
Point
Remark
Initial
Value
Disp. U / D
(*1)
C00 BASIC SETUP
C10 TAG NO. W 16 characters Tag Number
(*2)
C20 FLUID W LIQUID:Volume
GAS/STEAM:Volume
LIQUID:Mass
GAS/STEAM:Mass
GAS:STD/Normal
(0)
(1)
(2)
(3)
(4)
Selection of FLUID type (0)
(*2)
DL
(Indicate and Set only for C20 : LIQUID : Volume, GAS / STEAM: Volume)
C22 VOLUME UNIT W m
3
k m
3
l
cf
m cf
k cf
USgal
k USgal
UKgal
k UKgal
bbl
m bbl
k bbl
(0)
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
Selection of Flow Units for
Flow Rate
(0)
(*2)
DL
(Indicate and Set only for C20 : LIQUID : Mass, GAS / STEAM : Mass)
C25 DENSITY UNIT W kg/m
3
lb/cf
lb/USgal
lb/UKgal
(0)
(1)
(2)
(3)
Selection of Density Unit (0)
(*2)
DL
C26 DENSITY f W 0.00001 to 32000 C25 0 to 5 Operating Density
(Manual Setting Value)
1024
(*2)
DL
C27 MASS UNIT W kg
t
lb
k lb
(0)
(1)
(2)
(3)
Selection of Mass Flow Unit (0)
(*2)
DL
(Indicate and Set only for C20 : GAS : STD / Normal)
C30 TEMP UNIT W deg C
deg F
(0)
(1)
Selection of Temperature Unit (0)
(*2)
DL
C31
TEMP f
W -999.9 to 999.9
C30 1 Operating Temperature
(Manual Setting Value)
15.0
(*2)
DL
C32 TEMP b W -999.9 to 999.9 C30 1 Standard / Normal
Temperature
15.0
(*2)
DL
C33 PRESS UNIT W MPa abs
kPa abs
bar abs
kg/cm
2
a
psia
(0)
(1)
(2)
(3)
(4)
Selection of Pressure Unit (0)
(*2)
DL
C34 PRESS f W 0.00001 to 32000 C33 0 to 5 Absolute Pressure at
Operating Condition
(Manual Setting Value)
0.1013
(*2)
DL
C35 PRESS b W 0.00001 to 32000 C33 0 to 5 Absolute Pressure at Standard
Condition
0.1013
(*2)
DL
C36 DEVIATION W 0.001 to 10.0 3 Deviation Factor 1.0
(*2)
DL