Operation Manual
    <14. EXPLOSION PROTECTED TYPE INSTRUMENT>
14-12
IM 01F06A00-01EN
14.4 CSA
(1) Technical Data
• Explosion Proof
Applicable Standard: C22.1-98, C22.2 No.0-M1991, 
C22.2 No.0.4-04, C22.2 
No.0.5-1982, C22.2 No. 25-
1966, C22.2 No. 30-M1986, 
C22.2 No. 94-M1991, C22.2 
No. 142-M1987, C22.2 No. 
61010-1-04, ANSI/ISA-
12.27.01-2003
Certifi cate: 1166201
Type of Protection: 
Explosionproof for Class I, B, C and D; Class II, 
Groups E, F and G; Class III.
For Class I, Division 2 location:
“FACTORY SEALED, CONDUIT SEAL NOT 
REQUIRED.”
Enclosure : Type 4X
(Integral Type Vortex Flowmeter and Remote Type 
Vortex Flow Detector)
Temperature Code 
Ambient 
Temperature 
Process 
Temperature 
T6 60°C 85°C
T5 60°C 100°C
T4 60°C 135°C
T3 60°C 200°C
T2 60°C 300°C
T1 60°C 450°C 
Temperature Code: T6 (Remote Type Vortex Flow 
Converter)
Ambient Temperature: 
-29 to +60°C (Integral Type Vortex Flowmeter 
and Remote Type Vortex Flow Detector) 
-40 to +60°C (Remote Type Vortex Flow 
Converter)
Power Supply: 42Vdc max. (Integral Type Vortex 
Flowmeter and Remote Type Vortex 
Flow Converter)
Output Supply (Integral Type Vortex Flowmeter): 
Current Output; 4 to 20mAdc 
Pulse Output; On=2Vdc, 200mA 
Off=42Vdc, 4mA
Output Signal (Remote Type Vortex Flow Detector): 
Output Signal; 30Vp-p, 100Ap-p
Input/Output signal (Remote Type Vortex Flow 
Converter):   Current Output; 4 to 20mAdc 
Pulse; On=2Vdc, 20mA 
Off=42Vdc, 4mA 
  Input Signal; 30Vp-p, 100Ap-p
Electrical Connection: ANSI 1/2 female (Special)
• Intrinsically Safe
Type “n” and Non-incendive
Applicable Standard: C22.2 No. 0-M91, C22.2 
No. 0.4-2004, C22.2 No. 
157-M1987, C22.2 No. 213-
M1987, C22.2 No. 1010.1-92, 
CAN/CSA-E60079-0:02, 
CAN/CSA-E60079-11:02, 
CAN/CSA-E60079-15:02, 
ANSI/ ISA-12.27.01-2003
Certifi cate: 1198227
Type of Protection: 
Ex ia IIC T4...T1 and Ex nC IIC T4...T1
(Integral Type Vortex Flowmeter and Remote 
Type Vortex Flow Detector)
Ex ia IIC T4 and Ex nC IIC T4
(Remote Type Vortex Flow Converter)
(Integral Type Flowmeter and Remote Type Vortex 
Flow Detector)
Temperature Code 
Ambient 
Temperature 
Process 
Temperature 
T4 60°C 135°C
T3 60°C 200°C
T2 60°C 300°C
T1 60°C 450°C 
Ambient Temperature: 
-29 to +60°C (Integral Type Vortex Flowmeter and 
Remote Type Vortex Flow Detector)
-40 to +60°C (Remote Type Vortex Flow 
Converter)
Degree of Protection of Enclosure: IP67
Electrical Parameters: Ui=30Vdc, Ii=165mAdc, 
Pi=0.9W
  Ci=12nF, Li=0.15mH
Electrical Connection: ANSI 1/2 NPT female
Type of Protection: 
Intrinsically Safe for Class I,II,III, Div.1, Groups 
A,B,C,D,E,F and G, Non- incendive for Class I,II, 
Div.2, Groups A,B,C,D,E,F and G, Class III, Div.1
(Integral Type Vortex Flowmeter and Remote Type 
Vortex Flow Detector)
Temperature Code 
Ambient 
Temperature 
Process 
Temperature 
T4 60°C 135°C
T3 60°C 200°C
T2 60°C 300°C
T1 60°C 450°C 
Temperature Code: T4 (Remote Type Vortex Flow 
Converter)
Ambient Temperature: -40 to 60°C
Enclosure: Type 4X
Electrical Parameters: 
Vmax=30Vdc, 
Imax=165mAdc, Pmax=0.9W, 
Ci=12nF, Li=0.15mH