Table 3.1 (b) Straight pipe length and recommendations (2)
D: Nominal diameter (mm)
Description Figure
Fluid vibration:
For a gas line which uses a position-type or roots-type blower
compressor or a high-pressure liquid line (about 1MPa or more)
which uses piston-type or plunger-type pump, fluid vibrations may
be produced.
In these case, install valve on the upstream side of digitalYEWFLO.
For inevitable fluid vibration, put a vibration damping device such
as throttling plate or expansion section in the upstream side of
digitalYEWFLO.
digitalYEWFLO
digitalYEWFLO
Piston-type or plunger pump:
Install the accumulator on the upstream side of digitalYEWFLO to
reduce uid vibrations.
digitalYEWFLO
Valve positon (T-type piping exist):
When pulsation causes by a T-type piping exist, install the valve on
the upstream of the owmeter.
Example: As shown in the figure, when the valve V1 is turned off,
the fluid flow throught B as to meter A the flow is zero. But due to
the pulsating pressure is detected, the meter is zero point become
uctuating. To avoid this, change the valve V1 location to V1'.
Note: In case of the Reduced Bore Type, moisture may be
remained upstream of the owmeter. Drain it appropriately.
Relocating
Valve (Off)
A
B
Flow
V1’ V1
digitalYEWFLO
Pressure and Temperature Taps:
Pressure tap outlet: install this tap between 2D and 7D on the
downstream side of a owmeter.
Temperature tap outlet: install this on the downstream side 1D to
2D away from a pressure tap.
Pressure tap
Temperature tap
Upstream
Flow
downstream
2 to 7D 1 to 2D
digitalYEWFLO
Mounting Gasket:
Avoid mounting gaskets which protrude into the pipe line. This may
cause inaccurate readings.
Use the gaskets with bolt holes, even if digitalYEWFLO is the wafer
type.
When using a spiral gasket (without bolt holes), confirm the size
with the gasket -manufacturer, as standard items may not be used
for certain ange ratings.
Pipeline Flange
Pipeline
No good
digitalYEWFLO