Operation Manual

<13. GENERAL SPECIFICATIONS>
13-13
IM 01F06A00-01EN
13.5 Detailed Accuracy
Accuracy is the value in range of xed accuracy ow velocity. Refer to Table 13.7.
Volumetric ow rate at operation condition
Model Code Standard Type Multi-Variable Type (/MV) Reduced Bore Type (/R1) Reduced Bore Type (/R2)
Liquid
DY015
±1.0%
(20000Re<2000*D)
±0.75%
(2000*DRe)
DY025
±1.0%
(20000Re<1500*D)
±0.75%
(1500*DRe)
±1.0%
(20000Re<1500*D)
±0.75%
(1500*DRe)
±1.0%
DY040
±1.0%
(20000Re<1000*D)
±0.75%
(1000*DRe)
±1.0%
(20000Re<1000*D)
±0.75%
(1000*DRe)
±1.0%
DY050
DY080
DY100
DY150
±1.0%
(40000Re<1000*D)
±0.75%
(1000*DRe)
±1.0%
(40000Re<1000*D)
±0.75%
(1000*DRe)
DY200
±1.0%
(40000Re)
DY250
DY300
DY400
Gas,
Steam
DY015
±1.0%
(Velocity 35m/s or less)
±1.5%
(Velocity 35m/s to 80m/s)
DY025
±1.0%
(Velocity 35m/s or less)
±1.5%
(Velocity 35m/s to 80m/s)
±1.0%
(Velocity 35m/s or less)
±1.5%
(Velocity 35m/s to 80m/s)
DY040
±1.0%
(Velocity 35m/s or less)
±1.5%
(Velocity 35m/s to 80m/s)
DY050
DY080
DY100
DY150
DY200
DY250
DY300
DY400
D: Inner diameter of digitalYEWFLO (mm)
Re: Reynolds number (non unit)
Note: This table shows the accuracy of pulse output. In case of analog output, add up ± 0.1% of full scale to the values mentioned above.
Guarantee conditions of liquid volumetric ow rate: the accuracy of a product before shipment in our water actual test facility.
Totalized value of 2000 pulse or greater, straight pipe length: upper 10D or greater, lower 5D or greater, Fluid temp. 20 ± 10°C
Gas, Steam: The accuracy which is add up from liquid measurement accuracy.
The accuracy is con rmed by actual measured value of typical nominal size.