Operation Manual
<13. GENERAL SPECIFICATIONS>
13-3
IM 01F06A00-01EN
Status Output Function*:
Flow Switch:
In case fl ow rate decreases under the fl ow set
value,a status signal is output.
Status signal output mode can reverse (ON/
OFF).
Data Security During Power Failure:
Data (parameter, totalizer value, etc) storage
by EEPROM. No back-up battery required.
Correction:
Instrument Error Correction:
Vortex fl owmeter instrument errors can be
corrected by segment approximations.
Reynolds Number Correction:
Output error at Reynolds number 20000 or
less is corrected by using fi ve-break-point line-
segment approximation.
Gas Expansion Correction:
When measuring a compressibility gas and
steam, this expansion factor is useful to
correct the error at high velocity of fl ow (35m/s
or more).
Down-scale or Up-scale burn out.
In case a CPU or EEPROM failure occurs,
fl ow meter output the signal of Up-scale (21.6
mA or more).
Up-scale or Down-scale (3.6 mA or less) is
user-selectable through the fail mode alarm
jumper.
Indicator:
Flow rate (% or engineering units) or tem-
perature value and totalizer can be indicated
simultaneously.
Short message for self diagnostics indicated.
Local parameter setting can be operated by
key switches.
In mounting direction, the right and left 90° is
rotatable.
EMC Conformity Standards:
EN 61326-1: 2006 Class A, Table 2 (For use in
industrial locations), EN 61326-2-3: 2006
Note1: This instrument is a Class A product, and it is
designed for use in the industrial environment.
Please use this instrument in the industrial
environment only.
Note2: Use the metal conduit for the remote cable.
Pressure Equipment Directive:
Type of equipment: piping
Type of fl uid: liquid and gas
Group of fl uid: 1 and 2
Module: H
MODEL
DN(mm)*
PS(MPa)*
PS·DN(MPa·mm)
CATEGORY**
DY015 15 42 630 Article 3,
Paragraph 3
(SEP)
DY025 25 42 1050 Article 3,
Paragraph 3
(SEP)
DY040 40 42 1680 II
DY050 50 42 2100 II
DY080 80 42 3360 II
DY100 100 42 4200 II
DY150 150 42 6300 III
DY200 200 42 8400 III
DY250 250 42 10500 III
DY300 300 42 12600 III
DY400 400 25 10000 III
* PS: Maximum allowable pressure for Flow tube, DN: Nominal
size
** Refered to Table 6 coverd by ANNEX II of EC Directive on
Pressure Equipment Directive 97/23/EC
200
100
50
0
-50
-29
100
50
0
300
3URFHVV7HPSHUDWXUHÛ&
With Indicator
$PELHQW7HPSHUDWXUHÛ&
Without Indicator
85
-50
250
85
-29
80
80
F1301.ai
55
Figure 13.1 Ambient Temperature limit (Integal Type)