Operation Manual
    <9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>
9-9
IM 01F06A00-01EN
(3) Communication when set in the multidrop 
mode.
•  The HART confi guration tool searches for a 
device that is set in multidrop mode when it is 
turned on. When the HART confi guration tool 
is connected to the device, the polling address 
and the tag will be displayed.
•  Select the desired device. After that, normal 
communication to the selected device is 
possible. However, the communication speed 
will be slow.
(4)  Release the Multidrop Mode
To release multidrop mode, call up the Poll 
addr display and set the address to “0”.
Return Loop current mode to Enable.
9.10.5  Loop Test, Simulation, and Squawk
(1) Loop test
This feature can be used to output a fi xed current 
for loop checks.
•  Procedure to call up the Loop test (Method)
[Root Menu]  Diag/Service  Loop test
(2) Device Variable Simulation Function
(Effective only when setting to HART 7)
Using the simulation function, the output signal can 
be confi rmed by setting any value and status to the 
selected device variable.
Call up the parameter (Method) and follow the 
message shown.
After completing the step 5 in the next table, the 
simulation starts.
•  Procedure of device variable simulation
step1
Call up the 
parameter
[Root Menu]  Diag/Service  
Simulate (M)
2 Selection of 
Device Variable
Select one parameter from the list 
below
Off
Flow rate
Total
Temperature
Density
Percent range
Loop Current
3 Setting of Value Input the simulate value
4 Setting of Data 
quality
Select one parameter from the list 
below
Bad
Poor accuracy
Manual / Fixed
Good
5 Setting of Limit 
status
Select one parameter from the list 
below
Not limited
Low limited
High limited
Constant
NOTE
•  The simulations act on current, LCD display, 
communication and alarm.
•  The simulation of total fl ow rate acts on 
LCD display and communication, not on 
measuring total fl ow rate.
The measuring total fl ow rate is continuosly 
working during simulation.