Operation Manual
    <9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>
9-4
IM 01F06A00-01EN
Table 9.1  Burst parameters
Command parameter Burst Command
Burst Msg Trigger 
Mode
Burst Trigger Source Burst Trigger Units
PV
(fl ow rate)
Cmd1:PV Continuous — —
Window PV Depends on the 
assigned variable to 
PV
Rising
Falling
On-change
% range/current
(Percent of range, Loop current)
Cmd2: % range/
current
Continuous — —
Window % range %
Rising
Falling
On-change
Process vars/current
(Loop current, PV, SV, TV, QV)
Cmd3:Dyn vars/
current
Continuous — —
Window PV Depends on the 
assigned variable to 
PV
Rising
Falling
On-change
Process vars/%range/current with status
*1
(Select from fl ow rate, total fl ow, 
temperature, density, percent of range, 
loop current, PV, SV, TV, and QV)
*2
Cmd9:Device vars w/
Status
Continuous — —
Window Top of Burst Device 
Variables
Depends on maapping
Rising
Falling
On-change
Process vars/%range/current
(Select from fl ow rate, total fl ow, 
Temperature, Density, percent of range, 
loop current, PV, SV, TV, and QV)
*2
Cmd33:Device 
Variables
Continuous Top of Burst Device 
Variables 
Depends on the 
assigned variable to 
Burst Device Variable
Window
Rising
Falling
On-change
Self diagnosis information
Cmd48:Read 
Additional Device Status
Continuous — —
On-change All status —
*1: Output the data with time and status.
*2: Select at Burst Device Variables
Update time period of each measurement value is 
as follows;
•  Flow rate: Flow rate Update time period
•  Total: Total Update time period
•  Temperature: Temperature Update time period
•  Density: Density Update time period
9.10.2 Burst Mode
When the Burst mode is enabled, the device 
continuously sends up to three data listed in Table 
9.1. 
When the Burst mode is set to “Wired HART 
Enabled”, the device continuously sends alarm 
signal also. 
Refer to Subsection 9.10.3 “Event Notifi cation” for 
detail.
When changing the setting of Burst mode, set “Off” 
to the Burst mode. Default setting is “Off”.
(1) Burst Message
The parameters for Burst Message are as 
follows.
•  Transmit data: command parameters of Burst 
Command
• Device variables: 
• Update period
•  Transmit condition: choice in Burst Msg Trigger 
Mode
Refer to Table 9.1 for the combination between 
command parameter and transmit condition. 
[Root Menu]  Detailed setup  Confi gure 
outputs  HART output  Burst Condition  
Burst Message1, 2, 3  Burst command
NOTE
• In case of Burst Mode setting change, confi rm 
Burst Mode parameter is OFF.
• Prioritize to use the 'Burst Message 1'.