Operation Manual
    <6. PARAMETERS>
6-3
IM 01F06A00-01EN
(3)  Item C : BASIC SETUP
These items are for the basic parameters with setting before shipment.
The parameters, C20 to C50, are not indicated when option code “/MV” is selected and parameter item is 
selected in F10 except “Monitor only” or “Not use”.
A value in “( )” is the data corresponding to the indicator.
Item Name R / W Data Range Unit
Decimal
Point
Remark
Initial
Value
Disp. U / D 
(*1)
C00 BASIC SETUP
C10 TAG NO. W 16 characters Tag Number
(*2)
C20 FLUID W LIQUID:Volume 
GAS/STEAM:Volume
LIQUID:Mass 
GAS/STEAM:Mass 
GAS:STD/Normal
(0)
(1)
(2)
(3)
(4)
Selection of FLUID type (0) 
(*2)
DL
(Indicate and Set only for C20 : LIQUID : Volume, GAS / STEAM: Volume)
C22 VOLUME UNIT W m
3
k m
3
l 
cf 
m cf 
k cf 
USgal 
k USgal 
UKgal 
k UKgal 
bbl 
m bbl 
k bbl 
(0)
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
Selection of Flow Units for 
Flow Rate
(0)
(*2)
DL
(Indicate and Set only for C20 : LIQUID : Mass, GAS / STEAM : Mass)
C25 DENSITY UNIT W kg/m
3
lb/cf 
lb/USgal 
lb/UKgal 
(0)
(1)
(2)
(3)
Selection of Density Unit (0)
(*2)
DL
C26 DENSITY f W 0.00001 to 32000 C25 0 to 5 Operating Density 
(Manual Setting Value)
1024
(*2)
DL
C27 MASS UNIT W kg 
t 
lb 
k lb 
(0)
(1)
(2)
(3)
Selection of Mass Flow Unit (0) 
(*2)
DL
(Indicate and Set only for C20 : GAS : STD / Normal)
C30 TEMP UNIT W deg C 
deg F 
(0)
(1)
Selection of Temperature Unit (0) 
(*2)
DL
C31
TEMP f
W -999.9 to 999.9
C30 1 Operating Temperature 
(Manual Setting Value)
15.0 
(*2)
DL
C32 TEMP b W -999.9 to 999.9 C30 1 Standard / Normal 
Temperature
15.0 
(*2)
DL
C33 PRESS UNIT W MPa abs 
kPa abs 
bar abs 
kg/cm
2
 a 
psia 
(0)
(1)
(2)
(3)
(4)
Selection of Pressure Unit (0) 
(*2)
DL
C34 PRESS f W 0.00001 to 32000 C33 0 to 5 Absolute Pressure at 
Operating Condition
(Manual Setting Value)
0.1013 
(*2)
DL
C35 PRESS b W 0.00001 to 32000 C33 0 to 5 Absolute Pressure at Standard 
Condition
0.1013 
(*2)
DL
C36 DEVIATION W 0.001 to 10.0 3 Deviation Factor 1.0 
(*2)
DL