Operation Manual
    <12. TROUBLESHOOTING>
12-2
IM 01F06A00-01EN
12.3  No Output When The Fluid is Flowing
Is the output signal current
present?
Refer to Section 10.1 
"Adjustment" and loop test is OK?
No
Check the 
polarities.
Yes
Are parameters configured
correctly at operating 
conditions?
Yes
Yes
Is "K34:VORTEX FREQ" 0Hz?
Yes
Are the power polarities 
correct?
No
Are there any disconnected
power terminals?
Fasten the 
power terminal.
Yes
No
Is the cable broken?
Yes
Replace the 
cable.
No
Replace the 
AMP unit.
Replace the 
AMP unit.
No
Configure the 
parameters correctly.
No
Is the low cut configured correctly?
Configure the Low 
cut value correctly.
Is the TLA value configured 
correctly?
No
No
Configure the TLA value, refer to Section 
10.2 "Adjustment for Manual Mode."
No
Is the density value configured 
correctly?
Configure the density value correctly
at operating conditions.
No
Yes
Yes
Yes
Is output indicated when 
"K25:N.B MODE" is "MANUAL"
and "K26:NOISE RATIO" is "0"?
No
Is there broken the sensor?
No
Replace the 
AMP unit.
Replace the 
shedder bar
Yes
Yes
This case is due to coating of 
vortex shedder bar.
Check the shedder bar piping inner 
surfaces.
Yes
F1202.ai
Yes
Was a faulty area 
found with selfdiagnosis?
No
Check for recovery measures, refer to 
Section 6.5 "Self-Diagnostic (Error Code List)."