Instruction Manual

14-16
IM DR231-01E
14.2 Specifications of Optional Functions
GP-IB Communication (when equipped with /C1)
Electrical and mechanical specifications
conform to IEEE Standard 488-1978
Code
ISO (ASCII) code
Address
0 to 15
Functions
Talker functions
Output of measurement values (ASCII, binary); output of setting parameters (ASCII)
Listener functions
Setting of measurement conditions, controls of starting and stopping measurement, specifying
causes of interrupts (excluding setting and control of power ON/OFF)
Causes of interrupts
Syntax error, chart end, completion of A/D conversion, operations of internal timer, end of data
saving/reading, incomplete measurement during computation
RS-232-C Communication (when equipped with /C2)
Electrical and mechanical specifications
conform to standard EIA RS-232-C
Connection method
point-to-point
Communication method
half-duplex
Synchronization mode
Start-stop synchronization (synchronized by a start and a stop bit)
Baud rate
150, 300, 600, 1200, 2400, 4800, 9600, 19200, 38400 bps
Start bit
Fixed at 1 bit
Data length
7 or 8 bits, selectable
Parity
Selectable from even, odd, or none
Stop bit
1 or 2, selectable
Transmission distance
max. 15 m
Connector
D-sub 25pin
Handshake
hardware : transmission and reception control by DTR’’ RTS, CTS signal enabled.
Software : transmission control by XON and XOFF enabled.
Capacity of receiving buffer
200 bytes
Escape sequence
for reception only
Functions
Talker functions
Output of measurement data (ASCII, binary) and setting parameters (ASCII)
Listener functions
Setting of measurement conditions, control of measurement start and stop (excluding the setting
and control of power on/off), specifying causes of ESC S (output of a status byte)
Contents of status
Syntax error, chart end, completion of A/D conversion, operations of internal timer, end of data
saving/reading, incomplete measurement during computation