13.5 Calibration
Required Equipment
Name Measurement Accuracy Recommended
Range
DC Voltage Generator 0V to 50V 0.005% of setting+1µV Yokogawa 9000*, 4808
DMM 0V to 50V 0.005% of setting+1µV HP3458A
Decade Resistance Box 0.1Ω to 1000Ω 0.01% Yokogawa 2793
* For 0V input, it is necessary to either short the input terminals, or to monitor the output voltage using the
DMM.
Calibrating Conditions
Ambient temper : 23°C±2°C
Ambient humidity : 55 10%RH
AC power supply voltage : 100 to 240 VDC
DC power supply voltage : 12 to 28 VDC
Power supply frequency : 50/60 Hz±1% (for AC power supply model only)
Warm-up time : at least 30 minutes for this recorder, and necessary warm-up time
for the used equipment
Connection (terminal screw connections)
• DC Voltage Measurement
• Apply a rated voltage in the voltage range that is to be calibrated to channel 3.
+–
DC Voltage Generator
Input terminal
(CH3)
• Short-circuit between “+” and “–” terminals in channel 2.
+–
Input terminal
(CH2)
• Temperature Measurement Using RTD
• apply the 100Ω resistance to channel 5.
BbA
Input terminal
(CH4)
• Short-circuit the resistance input terminals A, B, and b in channel 4.
BbA
Decade Resistance Box
Input terminal
(CH5)