13.4 Error Codes
100 No equation option Not possible to make any computation
settings.
101 Invalid code in the equation Set the equation correctly.
102 Incorrect number of ( ) in the equation Set the correct number of ( ).
103 Syntax error Set the equation correctly.
104 Illegal character(s) or code(s) is used before Set the equation correctly.
or after an operator.
105 The right and left spans are the same. Change the span.
106 Attempted to make range setting during Stop making range setting.
computation.
110 Syntax error for computation constant Set the constant correctly.
111 Out of constant setting range Set the constant within the range.
120 Floppy disk not inserted or not formatted. Insert a formatted floppy disk into the
drive.
121 Insufficient memory Delete unnecessary files.
122 Incorrect file name or volume name Set the correct file name or volume name.
123 Attempted to make settings Stop making the settings.
which are not allowed to be made
during saving or reading of measured data.
124 Too many files in the media Delete unnecessary files.
125 Attempted to read data which was saved Select data which has been saved
by another instrument model. by DR/DA.
127 Attempted to save data in a write-protected file. Cancel write-protection.
128 No event/action has been set. Set the desired action .
130 Incorrect channel found during saving Set the correct channel.
or reading of measured data.
132 Attempted to format the disk while a file is open. Close the file.
134 Set-up data saving error
135 Internal data error
136 Incorrect copy channel Set the correct copy channel.
139 Other media related error
141 Disconnected modules were selected. Specify correct modules.
143 Modules selected which cannot be calibrated. Specify correct modules.
144 Calibration done improperly. Check electrical connections.
Calibrate again.
If the same error occurs again, contact
your nearest Sales representative.
145 Hardware becomes faulty during calibration. Turn off the power and then on.
Calibrate again.
If the same error occurs again, contact
your nearest Sales representative.
170 The IP address does not belong to any of the Set the correct IP address.
classes, A, B, or C.
171 There is a hole in the mask or the host address Set the correct subnet mask.
section is not released.
172 The net address section including the subnet Set the correct net address.
does not match the subnet section of the IP
address.
173 The host address section of the IP address is Set the correct IP address.
either all zeroes or all ones.
ROM ERROR System ROM error Contact your nearest sales
representative.
ROM ERROR* System ROM error Contact your nearest sales
representative.
RAM ERROR Main memory error Contact your nearest sales
representative.
RAM-DISK ERROR RAM disk error Contact your nearest sales
representative.
FLOPPY ERROR Floppy disk error Contact your nearest sales
representative.
MAIN NV READ ERROR Internal nonvolatile memory read error Contact your nearest sales
representative.
MAIN NV WRITE ERROR Internal nonvolatile memory write error Contact your nearest sales
representative.
CARRIAGE CANNOT MOVE Printer eror Contact your nearest sales
representative.
SYSTEM ERROR
System error ( : number) Check whether each unit or
(
: number) power code is connected
properly after turning off the
power switch. If the same
error occurs again even though
you turn on the power switch,
contact your nearest sales
representative.
RESET ERROR An error occurs between the main unit Same as above.
and communication module.