Instruction Manual

7-3
IM DR231-01E
Executing Recording
7
7.3 Starting Message Printing
D I S P M O D E
P R I N TF U N C
I N S D E L
R A N G E A L A R M
S E TC H A R T E N T E R
F E E D
M. F U N C 1
M. F U N C 2
RECORD
ALARM
CHART
KEYLOCK
R E C O R D
MSG_PRINT
>Select Function item
E S C
Press the FUNC key to enter the FUNC menu.
To display the print menu from the RANGE, ALARM, CHART or SET(SET3) menu, press the
DISP key.
Select / set
using the keys.
To escape from a lower menu, press the MODE (ESC) key. Its main menu will appear,
although new settings / selections will not be kept.
MSG PRINT
PRINT MESSAGE No.=01
ENTER
ENTER
(Main menu)
(Lower menu)
Starting message printing (MSG PRINT)
Select a message number from 01 to 20. For details on how to set the message, see Section 6.7,
“Entering Messages, Headers and Titles” on page 6-15.
While the instrument is recording, if the ENTER key is pressed, messages are printed.
If an alarm occurs, messages are also printed. See Section 6.5, “Setting the Alarm Printout” on
page 6-11.
Message printout buffer
Analog trend
Up to 10 message printouts can be stored in memory.
Information exceeding 10 message printouts will be discarded.
After one message printout, 9 are stored and another printout message can then be stored.
If 11 message printouts (exceeding the allowed number of 10) are entered, an asterisk (*)
will be printed at the top of the message when 10 message printouts are executed.
The message buffer clear function is available for canceling stored printout messages (see
page 9-17).
Logging mode
Up to 5 message printouts can be stored in memory.
Information exceeding 5 message printouts will be discarded.
If 6 message printouts (exceeding the allowed number of 5) are entered, an asterisk (*) will
be printed at the top of the message when 5 message printouts are executed.
The message buffer clear function is available for canceling stored printout messages (see