User Manual

IM 12J05D02-01E
10-2 Appendix
10-5. User setting table
FUNCTION SETTING DEFAULTS USER SETTINGS
Parameter specific functions
01 *S.TYPE 0 Galvanic (7 µA)
*I.CELL 0 3.75 µA
02 *CHECK 0.1.0 Zero Off
Slope On
Membrane Off
04 *SAL.TY 0 Off
Temperature measuring functions
10 *T.SENS 1 Pt1000
11 *T.UNIT 0 °C
12 *T.ADJ None
13 *T.MAN 0 Off
Calibration parameter functions
20 *∆T.SEC 60 sec.
*∆mg/l 0.05 mg/l
21 *0.CAL 0 Off
22 *ZERO 0 µA (nA)
*SLOPE 3.75 (50) µA (nA)
mA outputs
30 *mA 1.1 both 4-20mA
31 *OUTP.F 0.2 DO & Temp.
*D/R 0 Reverse (control)
32 *BURN 0.0 both off
33 *RG.mA2 5.0 mg/l only for PI control
34 *tI.mA2 100 sec only for PI control
35 *TABL1 21 pt table see code 31, §10-1
36 *TABL2 21 pt table see code 31, §10-1
37 *DAMP 0 sec.
38 *SWTCH 0 Off
Contacts
40 *S1 2.0.0 high process Al.
41 *S2 1.0.0 low process Al.
42 *S3 4.0.0 WASH
43 *S4 4.0.0 FAIL
44 *D.TIME 0.2 sec.
*P.HYST 0.1 mg/l
*T.HYST 0.5 °C
45 *RANGE 5.0 mg/l
*PER 10.0 sec.
*FREQ 70 p/min
46 *tI.CNT 100 sec. only for PI control
47 *EXPIR 0 off
*tE.min 15 min