User Manual
IM 12J05D02-01E
Parameter setting 5-27
Code Display Function Function detail X Y Z Default values
User Interface
50 *RET Auto return No return to meas. from HOLD 0 1
Return to meas after 10 min 1
51 *MODE add. to MAINT Setpoint adj. disabled 0 0.0.0
Setpoint adj. also in MAINT 1
Wash start disabled 0
Wash start in maintenance mode 1
Manual pressure disabled 0
Manual pressure enabled 1
52 *PASS Password Direct access to MAINT 0 0.0.0
MAINT protected by password #
Direct access to COMM 0
COMM protected by password #
Direct access to SERV 0
SERV protected by password #
53 *Err.01 Error setting Response & Time failure Soft fail 0
Response & Time failure Hard fail 1 1 Hard
*Err.02 Zero calibration Soft fail 0
Zero calibration Hard fail 1 1 Hard
*Err.03 Slope calibration Soft fail 0
Slope calibration Hard fail 1 1 Hard
*Err.04 Membrane failure Soft fail 0
Membrane failure Hard fail 1 1 Hard
*Err.07 Temperature sensor open Soft fail 0
Temperature sensor open Hard fail 1 1 Hard
*Err.08 Temp. sensor shorted Soft fail 0
Temp. sensor shorted Hard fail 1 1 Hard
*Err.09 Cell current abnormal Soft fail 0
Cell current abnormal Hard fail 1 1 Hard
*Err.12 Abnormal measured value Soft fail 0
Abnormal measured value Hard fail 1 1 Hard
*Err.16 Cell for maintenance Soft fail 0 0 Soft
interval time exceeded
Cell for maintenance Hard fail 1
interval time exceeded
*Err.22 Control time-out Soft fail 0 0 Soft
Control time-out Hard fail 1
*SOFT LCD Fail Constant 0 0 LCD Fail
LCD only 1
54 *E12 overrange limit E12 active at 0 - 300 % sat (Note) 150%
55 *CALL.M E16 setting No call for maintenance 0 0
Call for maint. enabled 1
E16 message after 1- 250 days 30
56 *UNIT PV unit Unit of measurement mg/l 0 0
Unit of measurement ppm 1
Unit of measurement % sat. 2
57-59 Not used
Pass: 111 for #=1,
333 for #=2,
777 for #=3,
888 for #=4,
123 for #=5,
957 for #=6,
331 for #=7,
546 for #=8,
847 for #=9
Note: For setting a value greater than 199.9, move the decimal point. For a value greater than 200%,
decimal places cannot be used.