Owner manual

<3. Installing the Sensor and Connecting the Cable>
3-3
IM 12J5B3-01E
[PB350GAngledFloatingBallHolder]
The tting is designed to contain the sensor in a sphere that oats on a liquid. This tting is im-
mune to large variations in the liquid level. Since the wet part is smooth and less susceptible to
catching ocs, the sensor does not trap rubbish that mixes in with the SUM (Related descrip-
tion: See Subsection 3.2.3). The oat is supported by an arm and does not move away from the
measurement point even if the SUM is owing.
When using this holder the membrane damage detector function in the converter can be acti-
vated.
Arm assembly
Sensor cable
Float
Mounting bracket
Fix the bracket to a nominal 50 A
pipe (OD of 60.5 mm) or a
horizontal surface (floor, the
edge of a solution tank, etc.).
F06.ai
Figure3.3 SuspensionofDissolvedOxygenSensorUsingPB350GAngledFloatingBallHolder
[DOX8HS(StyleB)ImmersionTypeHolder]
The holder holds the sensor at an angle of 45°. Consequently, bubbles will not stick to the perme-
able membranne. Any pollutants in the SUM can be automatically cleaned off at regular intervals
using a jet-spray water (air) cleaner.
Cleaning-utility pipe
Sensor cable
Dissolved oxygen sensor
Jet cleaner
F07.ai
Figure3.4 SuspensionofDissolvedOxygenSensorUsingDOX8HSImmersionTypeHolder
(withJetCleaner)
4th Edition : Apr 05,2011
1
2
3