Instruction Manual

Table Of Contents
4-10 Profibus
IM 12A00A01-61E
4-4-3-1. Transducer block parameters PH202 (continued)
Parameter Default Alternatives Unit R/W Slot, Index Data Type
(byte,bit) (bytes)
Reference impedance circuit
Low(0) High(1) R/W 4, 167 (2.3) unsigned32 (4)
Reference impedance
compensation Disabled(0) Enabled(1) R/W 4, 167 (2.4) unsigned32 (4)
Ref. impedance exceeds
low limit (E4.2) Hard fail(1) Soft fail(0) R/W 4, 166 (2.4) unsigned32 (4)
Ref. impedance exceeds
high limit (E5.2) Hard fail(1) Soft fail(0) R/W 4, 166 (2.5) unsigned32 (4)
Temp. sensor open (E7) Hard fail(1) Soft fail(0) R/W 4, 166 (2.6) unsigned32 (4)
Temp. sensor shorted (E8) Hard fail(1) Soft fail(0) R/W 4, 166 (2.7) unsigned32 (4)
Primary value exceeds
limits (E9) Hard fail(1) Soft fail(0) R/W 4, 166 (1.0) unsigned32 (4)
Calibration timer
expired (E16) Soft fail(0) Hard fail(1) R/W 4, 166 (1.7) unsigned32 (4)
Passcode Configuration
Passcode Maintenance - 000, 111, 333, 777, 888, 123, 957,
331, 546, 847 R/W 4, 160 unsigned16 (4)
Passcode Commissioning - 000, 111, 333, 777, 888, 123, 957,
331, 546, 847 R/W 4, 161 unsigned16 (4)
Passcode Service - 000, 111, 333, 777, 888, 123, 957,
331, 546, 847 R/W 4, 162 unsigned16 (4)
Display Settings
Display precision 0.1 pH(0) 0.01 pH(1) R/W 4, 167 (1.2) unsigned32 (4)
Auto return Enabled(1) Disabled(0) R/W 4, 167 (1.3) unsigned32 (4)
Logbook Configuration
power up Logbook 2(3), Not Logged(1), Logbook 1(2) R/W 4, 60 unsigned8 (1)
power down Logbook 2(3), Not Logged(1), Logbook 1(2) R/W 4, 61 unsigned8 (1)
system error Logbook 1(2), - R 4, 62 unsigned8 (1)
defaults loaded Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 63 unsigned8 (1)
logbook erased Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 64 unsigned8 (1)
init performed Logbook 1(2), - R 4, 65 unsigned8 (1)
error on Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 66 unsigned8 (1)
error off Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 67 unsigned8 (1)
temperature adjust Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 68 unsigned8 (1)
manual temperature Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 69 unsigned8 (1)
aspot changed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 70 unsigned8 (1)
slope changed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 71 unsigned8 (1)
aspot 2 changed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 72 unsigned8 (1)
temperature coefficient Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 73 unsigned8 (1)
input 1 imp after cal Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 74 unsigned8 (1)
input 2 imp after cal Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 75 unsigned8 (1)
delta ph changed Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 76 unsigned8 (1)
delta t changed Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 77 unsigned8 (1)
isothermal point changed Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 78 unsigned8 (1)
zeropoint changed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 79 unsigned8 (1)
new sensor installed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 80 unsigned8 (1)
one point calibration Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 81 unsigned8 (1)
two point calibration Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 82 unsigned8 (1)
Buffer Solution 1
Buffer1 ID 4.00 R/W 4, 105 unsigned8 (1)
Buffer1 at 0°C 4.00 R/W 4, 106 float (4)
Buffer1 at 5°C 4.00 R/W 4, 107 float (4)
Buffer1 at 10°C 4.00 R/W 4, 108 float (4)
Buffer1 at 15°C 4.00 R/W 4, 109 float (4)