< 3. INSTALLATION, PIPING AND WIRING >
3-4
IM 12T03A01-02E
Anchor bolt
Drain pit
Concrete base
Density meter sampling unit
Figure 3.6 Installation Example for Sampling Unit (Model VD6SM)
(2) Note of Designing a Density Meter Sampling Unit
Minimize liquid pipe length to improve response.
Design the piping not to allow air bubbles to remain in the pipe which may cause an error.
Piping should not have sharp bending where solids may accumulate.
Maintenance such as cleaning should be made easily.
Select element material most suitable for measuring liquid. Especially, for foodstuff
measurement, the selection should be done strictly.
(3) Installation of a Density Meter Sampling Unit
For the density detector with the sampling unit or with the sampling equipment user provided,
installing place must
be near to sample tap on the main pipe.
be free from vibration
meet the ambient conditions speci ed by the hazardous area (when the detector is model
VD6DF ameproof type).
Moreover, the sampling unit should be xed to a concrete base drained well.