Owner's manual

1-82 IM 04L31A01-01E
Control Operation Display > Message Summary
A list of written messages and the time the messages were written is displayed.
Up to 120 sets of message information can be stored to the internal memory. When
the number of message information exceeds 120, the information is overwritten from
the oldest information.
By scrolling the screen using up and down arrow keys, the message information in the
internal memory can be displayed.
By selecting a message from the list using up and down arrow keys, the historical
trend of the display data or event data containing the message can be recalled. For a
description on the historical trend display, see “
Control Operation Display > Historical
Trend
” in this section.
Number of the message displayed on the bottom line
Number of messages in the internal memory
Message string
Message string Date/time
when message as written
Name of the user who wrote
the message
Cursor
(Select a message using the cursor and select [INFORMATION] > [JUMP TO HISTORY]
on the display selection menu to display the data historical trend containing the
selected message.)
Control Operation Display > Memory Summary
The information pertaining to the display data and event data in the internal memory is
displayed.
By selecting the display data or event data using the arrow keys, the historical trend
display can be recalled. For a description on the historical trend display, see “
Control
Operation Display > Historical Trend
” in this section.
The number of manual sampled data, TLOG data (/M1 option), and report data (/M1
option) residing in the internal memory are displayed.
For models that have the alarm output relays (option), the ON/OFF status of the
relays are also listed.
Number of data points in the internal memory/maximum number
of data points that can be acquired in the internal memory
Date/time when the newest
data was acquired
Date/time when the data
acquisition started
Date/time when the data
acquisition ended*
Sampling count
Cursor
(Select the data using the cursor and select [INFORMATION] > [JUMP TO HISTORY]
on the display selection menu to display the historical trend of the selected data.)
Factor causing the end of data
acquisition
Data type (switch usin
left and right arrow keys)
• Display data
• Event data
* On models with the optional /BT1 batch header, the batch number and lot number for each file
can be displayed in place of the date and time the data acquisition ended.