User Manual

3-3
SM 04L31A01-01E
Adjustments
3
RTD Range
Control (1st Slot)
Number of Internal Group 1 Group 2
Control Loops 100 300 100 300
0 loops
2 loops CH1 CH2
Loops 4, 6 CH1 CH2 CH6 CH7
Monitor (2nd, 3rd Slot)
Model Group 1 Group 2
100 300 100 300
10 ch CH1 CH2
20 ch CH1 CH2 CH11 CH12
Calibration Procedure
The calibration procedure can be carried out using keys on the instrument or
communications commands.
Calibrating Using Keys
1. Turn ON the unit while holding down the UP arrow key. The calibration screen
appears.
2. Press the A/D Input Adjust soft key. The A/D adjust screen is displayed.
01-05, 06-10 shows control input, and 01-10, 11-20 shows measurement input.
3. Select a group to calibrate (01—05, 06—10, 01—10, 11—20).
3.1 Calibration of the Measuring Instrument's Input