Setting the Remote RJC Reference Channel (Valid Only When the Input Mode Is 
Set to RRJC (TC))
12.  Select the reference channel for the remote RJC from the Ref. Ch. list box under 
RRJC.
Note
If the reference channel setting for difference input or the reference channel setting for the 
remote RJC is inappropriate when switching from the Channel screen to the Display or Monitor 
screen, a Channel dialog box opens (see the figure below). To disable the measurement 
channels with inappropriate settings (clear the Monitor check box on the Channel screen) and 
move to the Display or Monitor screen, click Yes. To change the settings of the inappropriate 
measurement channels without moving to the next screen, click No.
Setting the Unit, Tag Number, and Tag Comment 
13.  Click the Unit, Tag No., and Tag Comment text box and enter the unit, tag number, 
and tag comment, respectively.
Setting the Time Constant of the First-Order Lag Filter (Valid Only When the 
Input Mode Is Set to Something Other than DI) 
14.  Select the time constant from the Filter list box.
Setting the Burnout (Valid Only When the Input Mode Is Set to TC)
15.  Select the direction in which the measured value is set off the range (+OVER or –
OVER) when a burnout occurs from the Burnout list box.
Setting the RJC (When the Input Mode Is Set to TC)
16.  Click the Type box under RJC and select Internal or External. 
17.  If you select External, click the Volt box under RJC and enter the compensation 
voltage.
Enter the tag comment
(up to 30 characters)
Enter the unit (up to six characters, only
when the Use check box of Scale is selected)
Enter the tag No.
(up to 15 characters)
Select the reference channel for the remote RJC
(valid only when the input mode is set to RRJC (TC))
Set the RJC
(valid only when the input
mode is set to TC)
Set the burnout
(valid only when the input mode
is set to TC)
Set the time constant of the first-order
lag filter
(when the input mode is other than DI)