Owner manual
4-6
IM WX103-01E
3.  After making the connections according to the explanation given on the calibration 
execution screen 1, click the Calibrate button.
  The gure below is an example of the 4-CH, High-Speed Universal Input Module.
Start calibration
To return to the Unit Information screen
To return to the previous screen
  The gure below is an example of the 8-CH, Medium-Speed Analog Output Module. 
Start calibration
To return to the Unit Information screen
To return to the previous screen
  When you click the Calibrate button, a message “Please wait and do not perform any 
other operation.” appears.
  When the calibration completes successfully, the calibration execution screen 2 
appears.
  If the calibration fails, a message “Failed to calibrate.” appears.
Note
For a description of the accuracy of the connected input module, the calibration accuracy, 
and the environmental conditions for calibration, see the MX100 Data Acquisition Unit User’s 
Manual.
4.2 Calibration Procedure